EVITA V600
Report
- Report Number
- 9611500-2022-00097
- Event Type
- Malfunction
- Date Received
- April 19, 2022
- Date of Event
- March 28, 2022
- Report Date
- June 7, 2022
- Manufacturer
- DRÄGERWERK AG & CO. KGAA
- Product Code
- QOV
- PMA / PMN Number
- EUA 200143
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INVESTIGATION HAS JUST STARTED; RESULTS WILL BE PROVIDED WITHIN A FOLLOW-UP REPORT.
THE AFFECTED EVITA V600 WAS CHECKED ON SITE BY DRÄGER SERVICE AND THE LOG FILE WAS SAVED FOR FURTHER ANALYSIS. THE REPORTED WARM STARTS OF THE VENTILATION UNIT IN THE MORNING OF (B)(6) 2022, COULD BE CONFIRMED BASED ON THE LOGBOOK ENTRIES. IN THE INITIAL ON-SITE ANALYSIS OF THE DEVICE, NO MALFUNCTION COULD BE IDENTIFIED. BASED ON THE LOGBOOK ENTRIES, THE PBA M48.3 INCLUDING THE RED AND BLUE USD CARDS WAS REPLACED AS A PREVENTIVE MEASURE. THE FUNCTION TEST AND RAM MEMORY STRESS TEST CARRIED OUT WITH THE CIRCUIT BOARD AND THE USD CARDS IN THE FOLLOWING DID NOT RESULT IN ANY DEVIATION FROM THE SPECIFICATION. ALL LOGGED WARM START PROCESSES WERE RELATED TO A PREVIOUS POWER FAILURE EVENT, THE EXACT CAUSE OF WHICH CANNOT BE DETERMINED CONCLUSIVELY FROM THE LOGBOOK. THE COURSE OF EVENTS INDICATED THAT THE DEVICE WAS REPEATEDLY SWITCHED ON AND OFF (CORRESPONDING TO THE POWER FAILURE EVENT) BY THE USER IN RESPONSE TO ANOTHER PERCEIVED MALFUNCTION. THE "DEVICE MALFUNCTION (14)" ALARM MESSAGE LOGGED BEFORE THE WARM START PROCESSES INDICATES A TEMPORARY DEVIATION IN THE DISPLAY THAT MAY HAVE CAUSED THE USER TO SWITCH THE DEVICE OFF AND ON AGAIN USING THE MAIN SWITCH. SINCE NO MALFUNCTION COULD BE FOUND DURING THE DEVICE ANALYSIS, THE CAUSE OF THE REPORTED DEVICE PROBLEM CANNOT BE CONCLUSIVELY IDENTIFIED. IF THE DEVICE DETECTS A DEVIATION OF THE VENTILATION UNIT, A WARM START OF THE VENTILATION UNIT AND CONTROL PANEL ARE INITIATED AT THE SAME TIME AS A SPECIFIED REACTION. DURING A WARM START, VENTILATION IS TEMPORARILY SUSPENDED AND AUXILIARY AUDIBLE ALARM SOUNDS. MEANWHILE, THE SAFETY VALVE IS OPENED TO AMBIENT TO ALLOW THE PATIENT TO BREATHE SPONTANEOUSLY. AFTER 8 SECONDS AT THE LATEST, VENTILATION IS AUTOMATICALLY RESUMED WITH UNCHANGED SETTINGS. THE WARM START OF THE CONTROL PANEL IS COMPLETED AFTER A MINUTE AT THE LATEST. MEANWHILE, THE USER CAN OBSERVE THE VENTILATION THAT HAS ALREADY BEEN RESUMED VIA THE OLED DISPLAY OF THE VENTILATION UNIT AND READ SAFETY-RELEVANT PARAMETERS SUCH AS FIO2 CONCENTRATION, MINUTE VOLUME AND AIRWAY PRESSURE. THE WARM START FINALLY IS INDICATED BY THE ALARM MESSAGE "VENTILATION UNIT RESTARTED" ON THE CONTROL PANEL WITH AN ACOUSTIC ALARM TONE SEQUENCE. THE NUMBER OF SIMILAR CASES IS WITHIN THE EXPECTED RANGE OF THE RESPECTIVE RISK ASSESSMENT AND THUS ACCEPTED.
IT WAS REPORTED THAT THE DEVICE PERFORMED WARMSTARTS DURING USE. NO PATIENT INJURY WAS REPORTED.
IT WAS REPORTED THAT THE DEVICE PERFORMED WARMSTARTS DURING USE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 52438 | EVITA V600 | VENTILATORS, INTENSIVE CARE | QOV | DRÄGERWERK AG & CO. KGAA | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |