FDA Adverse Event Malfunction Summary report: N

CHECK-IT

MDR report key: 14137609 · Received April 18, 2022

Report

Report Number
3016521623-2022-00158
Event Type
Malfunction
Date Received
April 18, 2022
Date of Event
April 11, 2022
Report Date
April 27, 2022
Manufacturer
LUCIRA HEALTH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT HAS NOT BEEN RETURNED. A DHR REVIEW OF THE ASSOCIATED KIT LOT NUMBER (K08A112101224M1) WAS COMPLETED, AND 0 DISCREPANCIES WERE FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT NOR ANY PREVIOUS COMPLAINTS FROM THIS CUSTOMER ASSOCIATED WITH "FALSE POSITIVE" RESULTS PRIOR TO RECEIVED DATE OF APRIL 11, 2022. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO FMEA-001 AND FMEA-004. THE COMPLAINT RATE FOR "FALSE POSITIVE" IS UNDER THE EXPECTED THRESHOLD OF 2.5% (LABEL CLAIM)/1% (INTERNAL WARNING LIMIT). LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A112101224M1: SAMPLE VIAL LOT DHR REVIEW: 2112124, 2112151, 2112297, 2112357, 2112363(ASSOCIATED INTERNAL LOT # 210762-2KA, 210762-2KB, 211346.1-1M, 2113461-2B, 2113461-2J). TEST LOT DHRS REVIEW: 2112221, 2112232, 2112247, 2112247, 2112267, 2112391, 2112405, 2112405, 2112391, 2112391 (ASSOCIATED INTERNAL LOT # 210762-26C, 210762-26L, 210762-26O, 210762-26P, 211346.2-1C, 2113462-4D, 2113462-3R, 2113462-3S2113462-3Y, 2113462-3Z). BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA (B)(4) CHECK-IT.

Additional Manufacturer Narrative · 0

PRODUCT HAS NOT BEEN RETURNED. A DHR REVIEW OF THE ASSOCIATED KIT LOT NUMBER (K08A112101224M1) WAS COMPLETED, AND 0 DISCREPANCIES WERE FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT NOR ANY PREVIOUS COMPLAINTS FROM THIS CUSTOMER ASSOCIATED WITH "FALSE POSITIVE" RESULTS PRIOR TO RECEIVED DATE OF APRIL 11, 2022. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO FMEA-001 AND FMEA-004. THE COMPLAINT RATE FOR "FALSE POSITIVE" IS UNDER THE EXPECTED THRESHOLD OF (B)(4)% (LABEL CLAIM)/1% (INTERNAL WARNING LIMIT). LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A112101224M1: SAMPLE VIAL LOT DHR REVIEW: 2112124, 2112151, 2112297, 2112357, 2112363(ASSOCIATED INTERNAL LOT # 210762-2KA, 210762-2KB, 211346.1-1M, 2113461-2B, 2113461-2J). TEST LOT DHRS REVIEW: 2112221, 2112232, 2112247, 2112247, 2112267, 2112391, 2112405, 2112405, 2112391, 2112391 (ASSOCIATED INTERNAL LOT # 210762-26C, 210762-26L, 210762-26O, 210762-26P, 211346.2-1C, 2113462-4D, 2113462-3R, 2113462-3S2113462-3Y, 2113462-3Z). A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA (B)(4) CHECK-IT.

Description of Event or Problem · 0

ONE DEVICE REPORTED AS HAVING A FALSE POSITIVE RESULT. THE COMPLAINANT PERFORMED A RAPID TEST WITH A NEGATIVE RESULT FOLLOWING THE POSITIVE RESULT FROM THE LUCIRA TEST KIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
658993 CHECK-IT COVID-19 TEST KIT QJR LUCIRA HEALTH, INC. K08A112101224M1

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other