FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG CARD

MDR report key: 14136352 · Received April 18, 2022

Report

Report Number
1221359-2022-01903
Event Type
Malfunction
Date Received
April 18, 2022
Date of Event
October 20, 2021
Report Date
April 18, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
10811877011290
PMA / PMN Number
EUA202537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: ABBOTT DIAGNOSTICS (B)(4) PERFORMED REVIEW OF THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING FOR KIT PART NUMBER 195-000 / LOT 131413 AND DEVICE PART NUMBER 195-430H / LOT 129460A. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 131413 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. RELATED MFR REPORT NUMBERS: 1221359-2022-01898 THROUGH 1221359-2022-01908.

Description of Event or Problem · 0

THE CUSTOMER REPORTED ELEVEN (11) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES IN (B)(6) 2021 ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT SIX (6) OF ELEVEN (11) AND LOT 131413 (TOTAL QUANTITY 1). THE PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2021 WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR TESTING WAS PERFORMED THE SAME DAY (PLATFORM: PERKINELMER) ON A NASAL WHICH GENERATED A NEGATIVE RESULT. THE PATIENT WAS ASYMPTOMATIC. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
819413 BINAXNOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 131413 10811877011290

Patients

Seq Age Sex Outcome Treatment
1 59 YR Female