FDA Adverse Event Other Summary report: N

PTS PANELS CHOL+HDL+GLU TEST STRIPS

MDR report key: 1412613 · Received July 7, 2009

Report

Report Number
1836135-2009-00002
Event Type
Other
Date Received
July 7, 2009
Date of Event
May 20, 2009
Report Date
July 7, 2009
Manufacturer
POLYMER TECHNOLOGY SYSTEMS, INC.
Product Code
NBW
PMA / PMN Number
K071507
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

A COMPLAINT WAS RECEIVED FROM A DISTRIBUTOR IN (B)(4) REGARDING UNDER RECOVERY OF CHOLESTEROL AND HDL CHOLESTEROL USING PTS PANELS/CARDIOCHEK CHOL+HDL+GLU PANEL TEST STRIPS, LOT I891. AN INTERNAL INVESTIGATION OF THIS COMPLAINT CONFIRMED THE UNDER RECOVERY OF THESE ANALYTES FOR BOTH THE RETURNED (PREVIOUSLY OPENED) VIALS AND QC RETAINED VIALS FROM THE SAME LOT. THE UNDER RECOVERY OF CHOLESTEROL AND HDL CHOLESTEROL WAS EVIDENT TO A SIMILAR MAGNITUDE AS REPORTED BY THE CUSTOMER. BASED ON THE SIMILAR LOSS OF REACTIVITY OF BOTH THE RETURNED MATERIAL AND THE RETAINED MATERIAL, THIS LOT OF CHOL+HDL+GLU IS BEING RECOVERED FROM THE FIELD. FURTHER EVAL IS NOW UNDERWAY TO DETERMINE THE ROOT CAUSE AND EVALUATE ANY PERFORMANCE IMPACT ON THE RECOVERY OF THE GLUCOSE ANALYTES IN THIS PANEL. WE HAVE SUBMITTED THIS MEDICAL DEVICE REPORT TO ALERT USERS OF THIS LOT NUMBER OF PRODUCT WHILE THE EVAL CONTINUES AND THE FIELD NOTIFICATION COMMENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PTS PANELS CHOL+HDL+GLU TEST STRIPS GLUCOSE AND LIPOPROTEIN TEST SYSTEMS NBW POLYMER TECHNOLOGY SYSTEMS, INC. NA I891

Patients

Seq Age Sex Outcome Treatment
1