FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 14121372 · Received April 15, 2022

Report

Report Number
1221359-2022-02234
Event Type
Malfunction
Date Received
April 15, 2022
Date of Event
March 12, 2022
Report Date
August 26, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 191-000 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 190-000. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORTS: 1221359-2022-02233 - 1221359-2022-02240.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1063017 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1063017, TEST BASE PART NUMBER 190-430 / LOT: 1063017. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1063017 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE; SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: THE REQUIRED INTAKE INFORMATION TO ENABLE FURTHER INVESTIGATION, SUCH AS THE KIT'S LOT NUMBER, WAS NOT PROVIDED AND AN INVESTIGATION WAS NOT ABLE TO BE PERFORMED. NOTWITHSTANDING, A REVIEW OF COMPLAINTS' TREND REVEALS THAT ALL LOTS WITHIN EXPIRY DATING ARE PERFORMING ACCORDING TO THE STATEMENTS MADE IN THE PACKAGE INSERT. IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUES AS NO INFORMATION WAS PROVIDED FOR INVESTIGATION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE IDNOW COVID-19 ASSAY FOR MULTIPLE PATIENTS WITH UNKNOWN LOT NUMBERS. THIS MFR. REPORT ADDRESSES PATIENT TWO (2) OF EIGHT (8). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE IDNOW COVID-19 ASSAY FOR PERFORMED ON (B)(6) R2022. CONFIRMATION TESTING (PLATFORM: CEPHID) WAS PERFORMED AND GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT WAS SYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1974995 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1063017 10811877011320

Patients

Seq Age Sex Outcome Treatment