FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 14119637 · Received April 15, 2022

Report

Report Number
1221359-2022-01881
Event Type
Malfunction
Date Received
April 15, 2022
Date of Event
March 22, 2022
Report Date
May 19, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
Product Code
QJR
UDI-DI
10811877011320
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS . A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORTS: 1221359-2022-01880 THROUGH 1221359-2022-01881.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED IN DUPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M174659 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 190-000 / LOT M174659 AND TEST BASE PART NUMBER 190-430 / LOT M174659. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M174659 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. REFERENCE MFR. REPORTS: 1221359-2022-01880 THROUGH 1221359-2022-01881.

Description of Event or Problem · 0

THE CUSTOMER REPORTED UNCONFIRMED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 (B)(6) 2022. REPEAT TESTING WAS PERFORMED WITH POSITIVE RESULTS.NO ADDITIONAL PATIENT INFORMATION,INCLUDING TREATMENT AND OUTCOME,WAS PROVIDED. THIS MFR. REPORT ADDRESSES TEST TWO OF TWO TESTS, LOT NUMBER (M174659) QUANTITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
138266 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC M174659 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 Unknown Congenital Anomaly