FDA Adverse Event Injury Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP

MDR report key: 14118705 · Received April 15, 2022

Report

Report Number
3007042319-2022-04709
Event Type
Injury
Date Received
April 15, 2022
Date of Event
March 3, 2022
Report Date
June 8, 2022
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION OF THIS EVENT IS PENDING AND A SUPPLEMENTAL REPORT WILL BE SENT UPON ITS COMPLETION.

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR INVESTIGATION COMPLETION. PRODUCT EVENT SUMMARY: THE VENTRICULAR ASSIST DEVICE (VAD) WAS NOT RETURNED FOR EVALUATION. THE REPORTED LOW FLOW EVENT WAS CONFIRMED VIA REVIEW OF THE AVAILABLE AUTO LOGS REPORTS, WHICH REVEALED SEVERAL DECREASES IN POWER CONSUMPTION AND ESTIMATED FLOWS STARTING 02/MAR/2022, WITH TRANSIENT DECREASES TO PARAMETERS BELOW NORMAL OPERATING RANGE, FOLLOWED BY AN INCREASE IN PARAMETERS STARTING ON 04/MAR/2022. ELEVEN (11) LOW FLOW ALARMS WERE LOGGED SINCE 03/MAR/2022. INFORMATION PROVIDED BY THE SITE INDICATED THAT, IN ADDITION TO LOW FLOWS, THE PATIENT PRESENTED WITH "FLAT" FREE HEMOGLOBIN (FHG) AND LACTATE DEHYDROGENASE (LDH). WHILE UNDER OBSERVATION, THE PATIENT'S RESPIRATORY CONDITION WORSENED DUE TO ASPIRATION PNEUMONIA, AND THE PATIENT WAS PLACED ON VENO-VENOUS EXTRACORPOREAL MEMBRANE OXYGENATION. THE PATIENT¿S ANTICOAGULATION WAS COMPLETELY REVERSED TO PERFORM SURGICAL TREATMENT FOR AN ILEUS AND INTRAVENOUS ANTICOAGULATION WAS STARTED. BASED ON THE AVAILABLE INFORMATION, THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE RISK DOCUMENTATION, POSSIBLE CAUSES OF THE REPORTED LOW FLOW EVENT MAY BE ATTRIBUTED TO MULTIPLE FACTORS INCLUDING BUT NOT LIMITED TO THROMBUS AT THE INFLOW CANNULA/OUTFLOW GRAFT, CONSTRICTION AT THE OUTFLOW GRAFT, AND/OR POOR VAD FILLING. PER THE INST RUCTIONS FOR USE, RESPIRATORY DYSFUNCTION IS A KNOWN POTENTIAL COMPLICATION ASSOCIATED WITH THE IMPLANTATION OF A VAD. BASED ON REVIEW OF PAST ADVERSE EVENTS FOR THIS PATIENT, IT WAS NOTED THAT THE PATIENT HAD A HISTORY OF RESPIRATORY DYSFUNCTION. POSSIBLE CLINICAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT INCLUDE THE PATIENT¿S PRE-EXISTING HISTORY AND RELATED COMORBIDITIES, THE PROGRESSION OF THEIR UNDERLYING DISEASE, ISSUES RELATED TO THE THERAPEUTIC USE OF ANTICOAGULANT AND ANTIPLATELET MEDICATIONS AND THE PATIENT'S COMPLEX POST-OPERATIVE COURSE. THERE ARE POSSIBLE PATIENT, PHARMACOLOGICAL AND PROCEDURAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT. INVESTIGATION OF THIS EVENT IS COMPLETED AND THE FILE WILL BE CLOSED. IF NEW INFORMATION IS RECEIVED, THE FILE WILL BE RE-OPENED AND A SUPPLEMENTAL WILL BE SUBMITTED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VENTRICULAR ASSIST DEVICE (VAD) PATIENT WAS HOSPITALIZED UNDER OBSERVATION DUE TO "FLAT" FREE HEMOGLOBIN (FHG) AND LACTATE DEHYDROGENASE (LDH). WHILE UNDER OBSERVATION, THE PATIENT'S RESPIRATORY CONDITION WORSENED DUE TO ASPIRATION PNEUMONIA, AND THE PATIENT WAS PLACED ON VENO-VENOUS EXTRACORPOREAL MEMBRANE OXYGENATION (V-V ECMO). THE PATIENT¿S ANTICOAGULATION WAS COMPLETELY REVERSED TO PERFORM SURGICAL TREATMENT FOR AN ILEUS AND INTRAVENOUS (IV) ANTICOAGULATION WAS STARTED. AFTER THE SURGERY, THE PATIENT¿S LACTATE DEHYDROGENASE (LDH) SLIGHTLY INCREASED AND THE VAD PUMP SPEED WAS INCREASED. IT WAS ALSO NOTED THAT THERE WERE MULTIPLE LOW FLOW ALARMS ON THE VAD. THE VAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174872 HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP VENTRICULAR (ASSISST) BYPASS DSQ HEARTWARE, INC. 1104

Patients

Seq Age Sex Outcome Treatment
1 58 YR Male Life Threatening| R| H