FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID 19 AG SELF TEST

MDR report key: 14118317 · Received April 15, 2022

Report

Report Number
1221359-2022-02199
Event Type
Malfunction
Date Received
April 15, 2022
Date of Event
November 8, 2021
Report Date
May 24, 2022
Manufacturer
ABBOTT DIAGNOSTIC SCARBOROUGH, INC
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 160538 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-160/195-260 / LOT 160538 AND DEVICE PART NUMBER 195-430H / LOT 150442. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE CURRENT OVERALL INCIDENT RATE FOR FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) FOR THIS SPECIFIC LOT BASED ON THE TOTAL QUANTITY OF DEVICES MANUFACTURED FOR DISTRIBUTION IS (B)(4). THE CURRENT OVERALL INCIDENT RATE FOR FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) FOR THIS SPECIFIC LOT BASED ON THE TOTAL QUANTITY OF DEVICES MANUFACTURED FOR DISTRIBUTION IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A CONFLICTING RESULTS RESULT WITH THE BINAXNOW COVID 19 AG SELF TEST (195-160) PERFORMED ON B)(6) 2021. THE CUSTOMER STATED THE PATIENT WAS SYMPTOMATIC (COLD AND FLU). NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2174067 BINAXNOW COVID 19 AG SELF TEST LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTIC SCARBOROUGH, INC 160538 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Female