FDA Adverse Event Injury Summary report: N

PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN)

MDR report key: 14117969 · Received April 15, 2022

Report

Report Number
1419937-2022-00040
Event Type
Injury
Date Received
April 15, 2022
Date of Event
March 29, 2022
Report Date
April 15, 2022
Manufacturer
MEDELA LLC
Product Code
HGX
UDI-DI
00020451401447
PMA / PMN Number
K200508
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER WAS SENT A REPLACEMENT PUMP AND RETURN OF HER ORIGINAL PUMP WAS REQUESTED FOR TESTING/EVALUATION. IN FOLLOW UP WITH A COMPLAINT HANDLER ON 4/7/2022, THE CUSTOMER INDICATED THAT THE REPLACEMENT PUMP WAS WORKING WELL AND THAT SHE WAS BETTER. THE DEVICE WAS RETURNED WITH THE CUSTOMER¿S POWER SUPPLY AND EVALUATED ON 4/12/2022. THE CUSTOMER¿S PUMP TURNED ON WITH THE CUSTOMER¿S POWER SUPPLY BUT POWERS OFF IF IT IS NOT HELD CORRECTLY. IT WAS DETERMINED IN THE EVALUATION THAT THE TRANSFORMER PLUG IS LOOSE IN THE DC JACK ON THE PUMP. THE CUSTOMER¿S REPORT OF A POWER ISSUE WAS CONFIRMED. NO PROBLEM FOUND WITH SUCTION WHEN TESTED. BASED ON THE RESULTS OF OUR INTERNAL INVESTIGATION (REFERENCE NUMBER (B)(4)),IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S MASTITIS. THE ESTIMATED INCIDENCE OF MASTITIS IN LACTATING WOMEN, WHETHER USING A BREAST PUMP OR NOT, ACCORDING TO PUBLISHED CLINICAL LITERATURE CAN BE AS HIGH AS 33%. IN FACT, CLINICAL GUIDELINES SUGGEST THE USE OF A BREAST PUMP TO FACILITATE WITHDRAWAL OF BREAST MILK DURING BOUTS OF MASTITIS. THE COMPLAINT RATE OF MASTITIS ACROSS ALL REPORTED FAILURES, ACROSS ALL MEDELA BREAST PUMPS, IS 0.008% FOR THE PERIOD OF JANUARY 2013 TO AUGUST 2017. MASTITIS IS USUALLY A BENIGN, SELF-LIMITING INFECTION WITH FEW CONSEQUENCES FOR THE SUCKLING INFANT. THE RISK OF MASTITIS IS HIGHER AMONG WOMEN WHO HAVE BREASTFED PREVIOUSLY, ESPECIALLY THOSE WITH A HISTORY OF MASTITIS. RIORDAN & WAMBACH, 4TH ED. P. 294: BREASTFEEDING AND HUMAN LACTATION. HOWEVER IN THIS CASE, THE MOTHER¿S MASTITIS REQUIRED PROMPT MEDICAL ATTENTION FOR PAIN RELIEF AND PRESCRIPTION ANTIBIOTICS TO AVOID PROGRESSION TO OVERWHELMING SEPSIS.

Description of Event or Problem · 0

ON (B)(6)2022, THE CUSTOMER ALLEGED TO MEDELA LLC THAT HER PUMP IN STYLE MAX FLOW HAS NOT BEEN WORKING WELL. THE CUSTOMER ALLEGED THAT THE BATTERY PACK WORKS WITH THE PUMP BUT THE POWER SUPPLY IS NOT WORKING. ADDITIONALLY, THE CUSTOMER ALLEGED THAT SHE WAS DIAGNOSED WITH MASTITIS ON(B)(6)2022 AND IS ON MEDICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1974796 PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN) PUMP, BREAST, POWERED HGX MEDELA LLC 101041360 0000020627 00020451401447

Patients

Seq Age Sex Outcome Treatment
1 Female Other