FDA Adverse Event Malfunction Summary report: N

DSD EDGE AUTOMATED ENDOSCOPE REPROCESSOR

MDR report key: 14117295 · Received April 15, 2022

Report

Report Number
2150060-2022-00023
Event Type
Malfunction
Date Received
April 15, 2022
Date of Event
March 16, 2022
Report Date
April 15, 2022
Manufacturer
MEDIVATORS, INC.
Product Code
FEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A MEDIVATORS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE UNIT AND FOUND THE GROUND FAULT INTERRUPTER (GFI) WAS CHARRED. THROUGH FURTHER INSPECTION, THE TECHNICIAN FOUND THAT THE UNIT WAS LEAKING FROM A CRACKED OVERFLOW COUPLING. THE GFI BECAME WET DUE TO THE LEAK RESULTING IN THE REPORTED EVENT. THE DSD EDGE HAS BEEN DESIGNED WITH FIRE-RESISTANT MATERIALS AND ANY THERMAL EVENTS WHICH OCCUR INTERNAL TO THE UNIT ARE EXPECTED TO BE CONTAINED AND NOT TO TRAVEL OUTSIDE OF THE UNIT. FURTHERMORE, THE SERVICE MANUAL PROVIDES THE FOLLOWING WARNING LANGUAGE FOR THE GFI AND POTENTIAL THERMAL EVENTS: "TO AVOID INJURY OR DEATH FROM AN ELECTRICAL INSULATION BREAKDOWN WITHIN THE UNIT, THE GFI (GROUND FAULT INTERRUPTER) CIRCUIT BREAKER SHOULD BE CHECKED FOR PROPER OPERATION ON AN ANNUAL BASIS. IF DURING THE USE OF THIS EQUIPMENT YOU SEE OR SMELL SMOKE, IMMEDIATELY DISCONNECT THE UNIT FROM THE POWER SUPPLY, DISCONTINUE USE AND CALL MEDIVATORS TECHNICAL SUPPORT." THE AER SUBJECT OF THIS REPORT WAS INSTALLED IN 2011 MAKING IT APPROXIMATELY 11 YEARS OLD. THE TECHNICIAN MADE THE NECESSARY REPAIRS, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem · 0

THE USER FACILITY REPORTED THAT SMOKE WAS EMITTING FROM INSIDE THEIR DSD EDGE AER. NO REPORT OF INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
140519 DSD EDGE AUTOMATED ENDOSCOPE REPROCESSOR ENDOSCOPE REPROCESSING SYSTEM FEB MEDIVATORS, INC.

Patients

Seq Age Sex Outcome Treatment
1 Unknown