DSD EDGE AUTOMATED ENDOSCOPE REPROCESSOR
Report
- Report Number
- 2150060-2022-00023
- Event Type
- Malfunction
- Date Received
- April 15, 2022
- Date of Event
- March 16, 2022
- Report Date
- April 15, 2022
- Manufacturer
- MEDIVATORS, INC.
- Product Code
- FEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A MEDIVATORS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE UNIT AND FOUND THE GROUND FAULT INTERRUPTER (GFI) WAS CHARRED. THROUGH FURTHER INSPECTION, THE TECHNICIAN FOUND THAT THE UNIT WAS LEAKING FROM A CRACKED OVERFLOW COUPLING. THE GFI BECAME WET DUE TO THE LEAK RESULTING IN THE REPORTED EVENT. THE DSD EDGE HAS BEEN DESIGNED WITH FIRE-RESISTANT MATERIALS AND ANY THERMAL EVENTS WHICH OCCUR INTERNAL TO THE UNIT ARE EXPECTED TO BE CONTAINED AND NOT TO TRAVEL OUTSIDE OF THE UNIT. FURTHERMORE, THE SERVICE MANUAL PROVIDES THE FOLLOWING WARNING LANGUAGE FOR THE GFI AND POTENTIAL THERMAL EVENTS: "TO AVOID INJURY OR DEATH FROM AN ELECTRICAL INSULATION BREAKDOWN WITHIN THE UNIT, THE GFI (GROUND FAULT INTERRUPTER) CIRCUIT BREAKER SHOULD BE CHECKED FOR PROPER OPERATION ON AN ANNUAL BASIS. IF DURING THE USE OF THIS EQUIPMENT YOU SEE OR SMELL SMOKE, IMMEDIATELY DISCONNECT THE UNIT FROM THE POWER SUPPLY, DISCONTINUE USE AND CALL MEDIVATORS TECHNICAL SUPPORT." THE AER SUBJECT OF THIS REPORT WAS INSTALLED IN 2011 MAKING IT APPROXIMATELY 11 YEARS OLD. THE TECHNICIAN MADE THE NECESSARY REPAIRS, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
THE USER FACILITY REPORTED THAT SMOKE WAS EMITTING FROM INSIDE THEIR DSD EDGE AER. NO REPORT OF INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 140519 | DSD EDGE AUTOMATED ENDOSCOPE REPROCESSOR | ENDOSCOPE REPROCESSING SYSTEM | FEB | MEDIVATORS, INC. |