MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Report
- Report Number
- 3006179046-2022-00172
- Event Type
- Malfunction
- Date Received
- April 14, 2022
- Report Date
- April 14, 2022
- Manufacturer
- NUVASIVE SPECIALIZED ORTHOPEDICS
- Product Code
- PGN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE IS UNABLE TO BE DETERMINED AT THIS TIME. IF ANY ADDITIONAL INFORMATION IS PROVIDED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. LITERATURE - TITLE OF ARTICLE: THE CROOKED ROD SIGN - A NEW RADIOLOGICAL SIGN TO DETECT DEFORMED THREADS IN THE DISTRACTION MECHANISM OF MAGNETICALLY CONTROLLED GROWING RODS AND A MODE OF DISTRACTION FAILURE - AUTHORS: JASON PUI YIN CHEUNG, TENG ZHANG, CORA BOW, KENNY KWAN, KAM YIN SZE, KENNETH MAN CHEE CHEUNG. PUBLICATION DATE: 2019.
INFORMATION WAS RECEIVED VIA LITERATURE THAT UPON REVIEW OF OTHER PATIENTS WHO HAVE EXPERIENCED DISTRACTION FAILURES AND CONTINUOUS ROD SLIPPAGE (OBSERVED IN 8 PATIENTS), NONE OBSERVED SIMILAR CROOKED ROD SIGN ON FOLLOW-UP RADIOGRAPHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 168163 | MAGEC SPINAL BRACING AND DISTRACTION SYSTEM | GROWING ROD SYSTEM - MAGNETIC ACTUATION | PGN | NUVASIVE SPECIALIZED ORTHOPEDICS | UNKNOWN | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |