T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2022-39566
- Event Type
- Malfunction
- Date Received
- April 14, 2022
- Date of Event
- March 18, 2022
- Report Date
- March 25, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PER TANDEM USER GUIDE: ALWAYS REMOVE ALL AIR BUBBLES FROM THE CARTRIDGE BEFORE BEGINNING INSULIN DELIVERY. ENSURE THERE ARE NO AIR BUBBLES WHEN DRAWING INSULIN INTO THE FILLING SYRINGE, HOLD THE PUMP WITH THE WHITE FILL PORT POINTED UP WHEN FILLING THE TUBING, AND ENSURE THAT THERE ARE NO AIR BUBBLES IN THE TUBING WHEN FILLING. AIR IN THE SYSTEM TAKES SPACE WHERE INSULIN SHOULD BE AND CAN AFFECT INSULIN DELIVERY. PER TANDEM USER GUIDE: AVOID EXPOSURE OF YOUR PUMP TO TEMPERATURES BELOW 41°F (5°C) OR ABOVE 99°F (37°C). INSULIN CAN FREEZE AT LOW TEMPERATURES OR DEGRADE AT HIGH TEMPERATURES. INSULIN THAT HAS BEEN EXPOSED TO CONDITIONS OUTSIDE OF THE MANUFACTURER¿S RECOMMENDED RANGES CAN AFFECT THE SAFETY AND PERFORMANCE OF THE PUMP.
IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. CUSTOMER CHANGED THE PUMP SUPPLIES AND SUCCESSFULLY RESUMED INSULIN THERAPY. REPORTEDLY, THE CUSTOMER IS REMOVING CARTRIDGE AIR BUBBLES BY DRAWING 250 UNITS, INSERTING NEEDLE INTO CARTRIDGE, PULLING UP ON PLUNGER TO PULL OUT AIR SEVERAL TIMES, REMOVING NEEDLE AND REMOVING AIR, THEN PUTTING BACK INTO CARTRIDGE, AND USING COLD INSULIN. CLINICAL SUPPORT ADVISED CUSTOMER THAT REMOVING CARTRIDGE AIR BUBBLES BY DRAWING 250 UNITS, INSERTING NEEDLE INTO CARTRIDGE, PULLING UP ON PLUNGER TO PULL OUT AIR SEVERAL TIMES, REMOVING NEEDLE AND REMOVING AIR, THEN PUTTING BACK INTO CARTRIDGE, AND USING COLD INSULIN, IS OFF LABEL PER THE TANDEM USER GUIDE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 250-259 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 133454 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Male | INFUSION SET: AUTOSOFT XCINSULIN: NOVOLOG |