BINAXNOW COVID-19 AG CARD
Report
- Report Number
- 1221359-2022-02034
- Event Type
- Malfunction
- Date Received
- April 13, 2022
- Date of Event
- January 21, 2022
- Report Date
- April 13, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 10811877011290
- PMA / PMN Number
- EUA202537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION: (B)(6). INVESTIGATION REPORT: ABBOTT DIAGNOSTICS (B)(4), INC. PERFORMED REVIEW OF MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING FOR KIT PART NUMBER 195-000 / LOT 165674 AND DEVICE PART NUMBER 195-430H / LOT 154520. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 165674 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. RELATED MFR REPORT NUMBERS: 1221359-2022-01916 THROUGH 1221359-2022-02048.
CUSTOMER REPORTED ONE HUNDRED THIRTY-THREE (133) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES IN (B)(6) ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT ONE HUNDRED NINETEEN (119) OF ONE HUNDRED THIRTY-THREE (133) AND LOT 165674 (TOTAL QUANTITY 7). PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2022. WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR NASAL SAMPLE WAS COLLECTED THE SAME DAY, PLACED IN VTM, AND TRANSPORTED VIA CURRIER TO A SEPARATE BRANCH LAB FOR TESTING (PLATFORM: PERKINELMER) WHICH GENERATED A NEGATIVE RESULT. THE PATIENT WAS ASYMPTOMATIC. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT. CUSTOMER IS PERFORMING TESTS IN COMMUNITY ORGANIZATIONS (COHORT 1: K-12 SCHOOLS INCLUDING FACULTY).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 900987 | BINAXNOW COVID-19 AG CARD | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 165674 | 10811877011290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Male | Hospitalization |