Narratives
FSE EVALUATED UNIT AND OBSERVED BATTERY 1 SHOWED 38650 MWH FOR 5 LEDS OF CHARGE, WHICH WOULD CORRELATE WITH THE BATTERY SHOWING LOW BATTERY IMMEDIATELY AFTER COMMENCING BATTERY RUNTIME. NOTHING UNUSUAL WAS IN THE LOGS. FSE REPLACED BATTERIES AS A PAIR AND COMPLETED PM. UNIT WAS CALIBRATED AND PASSED ALL FUNCTIONAL AND SAFETY TESTS PER FACTORY SPECIFICATIONS. RETURNED TO CUSTOMER AND CLEARED FOR CLINICAL USE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION. THE FULL EVENT SITE NAME IN IS (B)(6). THE INITIAL REPORTER NAMED IN IS A GETINGE EMPLOYEE WHO HAS DIFFERENT CONTACT DETAILS FROM THAT OF THE EVENT SITE. A CONTACT PERSON AT THE EVENT SITE IS (B)(6).
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE FIELD SERVICE ENGINEER (FSE) THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) BATTERY SHOWED FULL LEDS OF CHARGE WHEN UNIT WAS INITIALLY TESTED AND COMMENCED BATTERY RUNTIME TEST. AFTER ABOUT TWO MINUTES OF BATTERY RUNTIME BATTERY 1 SHOWED 3 LEDS OF CHARGE, ALARMED LOW BATTERY, AND LESS THAN 25 MINUTES OF RUNTIME ON THE SCREEN. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT WAS REPORTED.
N/A.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2563607 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | N/A | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | N/A. |