FDA Adverse Event Malfunction Summary report: N

RESPIRONICS

MDR report key: 14104529 · Received April 13, 2022

Report

Report Number
2031642-2022-00954
Event Type
Malfunction
Date Received
April 13, 2022
Date of Event
March 18, 2022
Manufacturer
RESPIRONICS CALIFORNIA, LLC
Product Code
MNT
UDI-DI
00884838020054
PMA / PMN Number
K102985
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PER BIOMED HE WAS USING CABLE, REMOTE ALARM, NORMALLY CLOSED (ALARM STATE = OPEN), (15.24 M) (50 FT) P/N 1003741 AND THIS REMOTE ALARM CABLE IS NOT TRIGGERING THE ALARM ON THE VENTILATOR. ADVISED BIOMED TO ORDER -CABLE, REMOTE ALARM, NORMALLY OPEN (ALARM STATE = CLOSED), (15.24 M) (50 FT) P/N 1003742 ,9898056118. PER BIOMED HE WILL ORDER THE CABLE AS INDICATED. NO FURTHER INFORMATION PROVIDED.

Additional Manufacturer Narrative · 0

PER THE BIOMEDICAL ENGINEER THEY WERE TESTING THE ALARM CABLES THAT THEY HAD STORED IN THEIR DEPARTMENT. THE CABLES WERE LABELED FOR THE V60. UNFORTUNATELY, THE CABLES DID NOT TRIGGER THE NURSE CALL SYSTEM APPROPRIATELY. THE BIOMEDICAL ENGINEER REPORTED THAT THE CABLES WERE NOT PLACED ON THE UNIT AND WERE NOT UTILIZED AT ALL. THE BIOMEDICAL ENGINEER REPORTED THAT THE CABLES FOR THE NURSE CALL SYSTEMS ARE CURRENTLY ON BACK ORDER. NO FURTHER INFORMATION PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER STATES THAT THE UNIT DOES NOT TRIGGER THE NURSE CALL EXCEPT WHEN IT IS FIRST TURNED ON. THE CUSTOMER REPORTED THAT THE UNIT WAS NOT IN USE ON A PATIENT. THE CUSTOMER CONTACTED PRODUCT SUPPORT AND IT WAS DETERMINED THAT THE CUSTOMER HAS A NORMALLY CLOSED CABLE. PRODUCT SUPPORT PROVIDED PART NUMBER FOR NORMALLY OPEN CABLE AND SUGGESTED TO CUSTOMER TO TRY THAT ONE. THE CUSTOMER WILL ORDER CORRECT NURSE CALL CABLE. FURTHER INFORMATION HAS BEEN REQUESTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AT A LATER DATE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2564638 RESPIRONICS VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE MNT RESPIRONICS CALIFORNIA, LLC V60 00884838020054

Patients

Seq Age Sex Outcome Treatment
1 Unknown