BINAXNOW COVID-19 AG CARD
Report
- Report Number
- 1221359-2022-02019
- Event Type
- Malfunction
- Date Received
- April 13, 2022
- Date of Event
- January 18, 2022
- Report Date
- April 13, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 10811877011290
- PMA / PMN Number
- EUA202537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
01471749-104. ABBOTT DIAGNOSTICS (B)(4), INC. PERFORMED REVIEW OF MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING FOR KIT PART NUMBER 195-000 / LOT 164258 AND DEVICE PART NUMBER 195-430WL / LOT 154627. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 155933 SHOWED THAT THE COMPLAINT RATE IS (B)(4)%. ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. RELATED MFR REPORT NUMBERS: 1221359-2022-01916 THROUGH 1221359-2022-02048.
CUSTOMER REPORTED ONE HUNDRED THIRTY-THREE (133) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES IN JANUARY ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT ONE HUNDRED FOUR (104) OF ONE HUNDRED THIRTY-THREE (133) AND LOT 164258 (TOTAL QUANTITY 10). PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON 18JAN2022 WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR NASAL SAMPLE WAS COLLECTED THE SAME DAY, PLACED IN VTM, AND TRANSPORTED VIA CURRIER TO A SEPARATE BRANCH LAB FOR TESTING (PLATFORM: PERKINELMER) WHICH GENERATED A NEGATIVE RESULT. THE PATIENT WAS ASYMPTOMATIC. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT. CUSTOMER IS PERFORMING TESTS IN COMMUNITY ORGANIZATIONS (COHORT 1: K-12 SCHOOLS INCLUDING FACULTY).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1545348 | BINAXNOW COVID-19 AG CARD | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 164258 | 10811877011290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Male | Hospitalization |