THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Report
- Report Number
- 2029046-2022-00781
- Event Type
- Injury
- Date Received
- April 13, 2022
- Date of Event
- March 18, 2022
- Report Date
- May 4, 2022
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- LPB
- UDI-DI
- 10846835010183
- PMA / PMN Number
- P030031
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE BWI PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION ON 01-APR-2022. THE DEVICE EVALUATION WAS COMPLETED ON 12-APR-2022. THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER FOR EVALUATION. A VISUAL INSPECTION AND AN EVALUATION OF ALL FEATURES OF THE DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED NO DAMAGE OR ANOMALIES ON THE DEVICE. PER THE EVENT, SEVERAL TESTS WERE PERFORMED. THE MAGNETIC, TEMPERATURE AND FORCE FEATURES WERE TESTED AND NO ISSUES WERE OBSERVED. IN ADDITION, THE DEVICE WAS DEFLECTING AND IRRIGATING CORRECTLY. NO MALFUNCTIONS WERE OBSERVED DURING THE ANALYSIS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER, AND NO INTERNAL ACTIONS WERE IDENTIFIED. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE IFU (INSTRUCTION FOR USE) STATES THAT CAREFUL CATHETER MANIPULATION MUST BE PERFORMED TO AVOID CARDIAC DAMAGE, PERFORATION OR TAMPONADE. THE ROOT CAUSE OF THE ADVERSE EVENT REMAINS UNKNOWN. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT COULD NOT BE REPLICATED DURING THE ANALYSIS. BIOSENSE WEBSTER MANUFACTURER'S REFERENCE NUMBER (B)(4) HAS TWO COMPLAINTS THAT ARE RELATED TO THE SAME INCIDENT. MANUFACTURER'S REFERENCE NUMBER: (B)(4)
ADDITIONAL INFORMATION WAS RECEIVED ON THE EVENT ON 11-APR-2022. THE CATHETER THAT WAS KINKED WAS THE ABBOTT INQUIRY. THE PHYSICIAN DID NOT SPECIFY IF THE ISSUE WAS CAUSED BY THE ABBOTT OR THE WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY. WHEN THE PHYSICIAN WAS ASKED DURING THE FOLLOW UP, HE ONLY MENTIONED THE TRAUMA WITHIN THE CORONARY SINUS WAS LIKELY ¿DUE TO THE CATHETER MANIPULATION WITHIN THE CORONARY SINUS¿ AND THAT IS ALL HE SPECIFIED. THE WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY DID NOT HAVE A KINK IN IT, WHILE THE OTHER ABBOTT INQUIRY CS CATHETER DID. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4)
IT WAS REPORTED THAT A (B)(6) MALE UNDERWENT A SUPRAVENTRICULAR TACHYCARDIA (SVT) ABLATION PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER AND WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY. THE PATIENT SUFFERED CARDIAC TAMPONADE REQUIRING A PERICARDIOCENTESIS. A CARDIAC TAMPONADE WAS NOTICED WHEN THE PATIENT¿S BLOOD PRESSURE (BP) AND HEART RATE (HR) DROPPED AFTER PACING. THE CARDIAC TAMPONADE WAS DIAGNOSED VIA INTRA CARDIAC ECHO (ICE) CATHETER. A PERICARDIOCENTESIS WAS PERFORMED BY THE PHYSICIAN AND THE PATIENT WAS STABLE AT THE TIME OF THE CALL. THE PHYSICIAN HAD ISSUES ADVANCING THE WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY AT THE BEGINNING OF THE PROCEDURE. THE CATHETER WAS EXCHANGED FOR AN INQUIRY (ABBOTT) CATHETER WHICH ENDED UP KINKED NEAR THE IVC (SEEN ON FLUORO). THE PHYSICIAN BELIEVES THAT CATHETER MANIPULATION WITHIN THE WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY MAY HAVE BEEN THE CAUSE. THE ADVERSE EVENT WAS DISCOVERED DURING USE BIOSENSE WEBSTER PRODUCTS. POST PERICARDIOCENTESIS, THE PATIENT STABILIZED AND THE PATIENT WAS REMOVED FROM THE ROOM. THE GENERATOR USED WAS A SMARTABLATE GENERATOR KIT-US. A TRANSSEPTAL PUNCTURE WAS PERFORMED WITH A VERSACROSS SYSTEM. THERE WAS NO KNOWN ISSUE WITH THE WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY. IT WAS BRIEFLY USED BEFORE BEING REPLACED WITH AN INQUIRY CS CATHETER. PRIOR TO NOTING THE CARDIAC TAMPONADE ABLATION WAS PERFORMED AND THERE WAS NO EVIDENCE OF A STEAM POP. IF IT DID OCCUR WITH CS MANIPULATION, IT WOULD HAVE BEEN AT THE BEGINNING OF THE CASE. THIS WOULD MAKE THE MOST SENSE AS THE PATIENT STEADILY STRUGGLED MORE AND MORE TO MAINTAIN BLOOD PRESSURE THROUGHOUT THE PROCEDURE. THE CARDIAC TAMPONADE WAS DISCOVERED DURING ABLATION PHASE OF THE PROCEDURE. AN IRRIGATED CATHETER WAS USED AND THE SMART TOUCH SF FLOW RATE WAS 2ML/MIN IDLE FLOW, 8ML/MIN <30 WATTS LOW FLOW & 15ML/MIN >30 WATTS HIGH FLOW. THE CORRECT CATHETER SETTINGS WERE SELECTED ON THE GENERATOR AND THE PUMP WAS SWITCHING FROM LOW TO HIGH FLOW DURING ABLATION. NO ERROR MESSAGES OBSERVED ON BIOSENSE WEBSTER EQUIPMENT DURING THE PROCEDURE. FORCE VISUALIZATION USED WAS GRAPH WITH THE VISITAG MODULE PARAMETERS FOR STABILITY SET TO 2-3MM, 3SEC, 25%, 3MM WITH NO ADDITIONAL FILTER USED. BASED ON THE AVAILABLE INFORMATION, THE THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER AND WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY WILL BOTH BE REPORTABLE AS ABLATION WAS PERFORMED AND THE PHYSICIAN FELT THAT CATHETER MANIPULATION WITHIN THE CORONARY SINUS MAY HAVE BEEN THE CAUSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 940706 | THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER | CARDIAC ABLATION PERCUTANEOUS CATHETER | LPB | BIOSENSE WEBSTER INC | D134805 | 30689751L | 10846835010183 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR | Male | Life Threatening| R | NON BWI-ABBOTT INQUIRY CS CATHETER| NON BWI-VERSACROSS SYSTEM| SMARTABLATE GENERATOR KIT-US| UNKNOWN BRAND ICE CATHETER| UNKNOWN BRAND PUMP| UNK_CARTO 3| UNK_WEBSTER CS WITH AUTO ID |