IMMULITE 2000
Report
- Report Number
- 2247117-2009-00019
- Event Type
- Other
- Date Received
- June 8, 2009
- Date of Event
- April 22, 2009
- Report Date
- May 27, 2009
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- DGC
- PMA / PMN Number
- K970227
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENTS SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
USER ERROR: A SIEMENS HEALTHCARE TECHNICAL SUPPORT ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA AND DETERMINED THAT THE CAUSE OF THE DISCORDANT RESULTS WERE DUE TO THE OPERATOR RUNNING THE INSTRUMENT WITHOUT THE REQUIRED WATER SUPPLY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT MULTIPLE ASSAY RESULTS WERE OBTAINED ON MULTIPLE PATIENT'S SAMPLES. THE SAMPLES WERE REPEATED AND THE CORRECT RESULTS WERE REPORTED. PATIENT TREATMENT WAS PRESCRIBED FOR TWO PATIENTS. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMULITE 2000 | IMMUNO-ASSAY ANALYZER | DGC | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ||||
| 10 | ||||
| 11 | ||||
| 12 | ||||
| 13 | ||||
| 14 | ||||
| 15 | ||||
| 16 | ||||
| 17 | ||||
| 18 | ||||
| 19 | ||||
| 2 | ||||
| 20 | ||||
| 21 | ||||
| 22 | ||||
| 23 | ||||
| 24 | ||||
| 25 | ||||
| 26 | ||||
| 27 | ||||
| 28 | ||||
| 29 | ||||
| 3 | ||||
| 30 | ||||
| 31 | ||||
| 32 | ||||
| 33 | ||||
| 34 | ||||
| 35 | ||||
| 36 | ||||
| 37 | ||||
| 38 | ||||
| 39 | ||||
| 4 | ||||
| 40 | ||||
| 41 | ||||
| 42 | ||||
| 43 | ||||
| 44 | ||||
| 45 | ||||
| 5 | ||||
| 6 | ||||
| 7 | ||||
| 8 | ||||
| 9 |