FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,11.5

MDR report key: 14086527 · Received April 12, 2022

Report

Report Number
0002023141-2022-00894
Event Type
Injury
Date Received
April 12, 2022
Date of Event
March 30, 2022
Report Date
September 28, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019232
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-00895. ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. FAX NUMBER UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-00895-1. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBERS (1251196 AND 1238711). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP150 AND OP150) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS (1251196 AND 1238711) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATIONS, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATIONS, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENTS REMAIN NON-VERIFIABLE AS THEY ARE A MEDICAL CONDITION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH PERI-IMPLANTITIS ON BOTH IMPLANTS. TOOTH SITES #24-25.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1322848 IMP,TSV,4.1MM,SBM,11.5 DENTAL IMPLANT DZE ZIMMER DENTAL TSV4B11 1251196 00889024019232

Patients

Seq Age Sex Outcome Treatment
1 64 YR Female Required Intervention