FDA Adverse Event
Malfunction
Summary report: N
PORTABLE THERAPEUTIC LIQUID OXYGEN SYSTEM
MDR report key: 14073321
·
Received April 11, 2022
Report
- Report Number
- 14073321
- Event Type
- Malfunction
- Date Received
- April 11, 2022
- Date of Event
- February 27, 2022
- Report Date
- April 6, 2022
- Manufacturer
- ESSEX INDUSTRIES, INC.
- Product Code
- BYJ
- UDI-DI
- 10815363020025
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
THE LIQUID OXYGEN SYSTEM FAILED, RAN COMPLETELY OUT OF OXYGEN 90 MINUTES INTO A PATIENT FIXED WING TRANSPORT WITH 90 PLUS MINUTES OF THE TRANSPORT REMAINING. THE PATIENT WAS ON ECMO, SO BOTH THE ECMO DEVICE AND VENTILATOR WERE USING OXYGEN. THE CREW HAD TO DIVERT THE FLIGHT TO THE CLOSEST ECMO CAPABLE CENTER AND MANAGE THE PATIENT ON A PORTABLE OXYGEN CYLINDER ON THE AIRCRAFT UNTIL MET AT THE AIRPORT. THE ESSEX INDUSTRIES, INC 10 LITER LIQUID OXYGEN CONVERTOR SYSTEM HAD A PRESSURE RELEASE VALVE/VENT FAILURE SOMETIME DURING THE COURSE OF THE TRANSPORT AND LOST ALL OXYGEN. SENT BACK TO MFG. FOR REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2445072 | PORTABLE THERAPEUTIC LIQUID OXYGEN SYSTEM | UNIT, LIQUID-OXYGEN, PORTABLE | BYJ | ESSEX INDUSTRIES, INC. | 50C-0021-1 | 10815363020025 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |