FDA Adverse Event Other Summary report: N

ROLLATOR, SUPER LIGHT, 250LB CAP

MDR report key: 1406872 · Received June 11, 2009

Report

Report Number
1417592-2009-00035
Event Type
Other
Date Received
June 11, 2009
Report Date
June 5, 2009
Manufacturer
MEDLINE INDUSTRIES, INC.
Product Code
ITJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT THE END USER SAT DOWN ON THE ROLLATOR AND ONE BRAKE DID NOT LOCK. SHE FELL AND FRACTURED A COUPLE OF RIBS. SHE WAS REPORTED TO BE DOING WELL. THE SAMPLE WAS RETURNED AND EVALUATED. THE FRONT AND REAR LEG EXTENSIONS WERE NOT SET EVENLY. THE REAR EXTENSIONS WERE SET AT ADJUSTMENT HOLE 1, AND THE FRONT EXTENSIONS WERE SET AT ADJUSTMENT HOLE 2. SINCE THE FRONT EXTENSIONS WERE SET HIGHER, IT GIVES THE SEAT A DOWNWARD SLOPING ANGLE. IT WAS CONFIRMED THAT THE LEFT BRAKE LOCKED AND THE RIGHT BRAKE DID NOT LOCK. WITH ADJUSTMENT, THE RIGHT BRAKE LOCKED AS SPECIFIED BY THE MANUFACTURER. INSTRUCTIONS FOR SET UP AND DEVICE MAINTENANCE ARE DESCRIBED IN THE OWNER'S MANUAL. THIS IS NOT A MANUFACTURING DEFECT AND THE INCIDENT APPEARS TO BE CAUSED BY A LACK OF MAINTENANCE. HOWEVER, DUE TO THE REPORT OF FRACTURED RIBS, THIS MDR IS BEING FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE END USER SAT DOWN ON THE ROLLATOR AND THE BRAKE DID NOT LOCK. SHE FELL, FRACTURING SOME RIBS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROLLATOR, SUPER LIGHT, 250LB CAP NONE ITJ MEDLINE INDUSTRIES, INC. MDS86825SLR A080926942

Patients

Seq Age Sex Outcome Treatment
1 Other