FDA Adverse Event Malfunction Summary report: N

BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD

MDR report key: 14056060 · Received April 8, 2022

Report

Report Number
9680577-2022-00053
Event Type
Malfunction
Date Received
April 8, 2022
Date of Event
March 16, 2022
Report Date
April 28, 2022
Manufacturer
BECTON DICKINSON GMBH
Product Code
JSJ
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD. CATALOG NUMBER 221353 WITH 510K NUMBER PREAMENDMENT. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY IT WAS REPORTED THAT 54 PLATES WOULD HAVE SHOW CONTAMINATION. THE COMPLAINT TRENDS WERE REVIEWED AND SIMILAR COMPLAINTS WERE FOUND FOR THIS LOT NUMBER. THEREFORE, A MRB (MATERIAL REVIEW BOARD) WAS PERFORMED AND THIS ISSUE WILL BE INVESTIGATED WITHIN A CAPA (CORRECTIVE AND PREVENTIVE ACTION) 4344556. THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. PICTURES WERE PROVIDED SHOWING THE REPORTED CONTAMINATION. THE RETENTION SAMPLES WERE REVIEWED AND THE PLATES WERE FOUND TO BE WITHOUT DEVIATION. AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. PLEASE NOTE THAT THIS PRODUCT DOES NOT HAVE AN SAL CLAIM. IT IS FILLED ASEPTICALLY, THEREFORE AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. WE ONLY RELEASE PRODUCT BATCHES TO THE MARKET WITH AN AQL (ACCEPTABLE QUALITY LEVEL) = 1.5. ALTHOUGH THIS CONTAMINATION LEVEL IS VERY LOW, WE CANNOT COMPLETELY EXCLUDE THAT A CUSTOMER MAY RECEIVE A CONTAMINATED PLATE. AS SEVERAL COMPLAINTS WERE REPORTED FOR THIS PRODUCT LOT NUMBER REGARDING CONTAMINATION, A MRB (MRB 22-006) AND CAPA (# 4344556) WERE CREATED FOR FURTHER EVALUATION. THE ROOT CAUSE INVESTIGATION IS CURRENTLY ONGOING WITHIN CAPA 4344556. BASED UPON OUR INVESTIGATION THE COMPLAINT WAS CONFIRMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WERE 54 BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD THAT HAD VISIBLE CONTAMINATION ON THE AGAR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE LAST DELIVERY WE RECEIVED FROM YOU OF COLUMBIA CNA AGAR WITH 5% SB, IMPROVED IS LARGELY CONTAMINATED. EVERY 10 PLATES WE HAVE TO DISCARD 3-6 BECAUSE THERE IS SOME FORM OF SIDE CONTAMINATION VISIBLE ON THE AGAR.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WERE 54 BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD THAT HAD VISIBLE CONTAMINATION ON THE AGAR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE LAST DELIVERY WE RECEIVED FROM YOU OF COLUMBIA CNA AGAR WITH 5% SB, IMPROVED IS LARGELY CONTAMINATED. EVERY 10 PLATES WE HAVE TO DISCARD 3-6 BECAUSE THERE IS SOME FORM OF SIDE CONTAMINATION VISIBLE ON THE AGAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1300425 BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL JSJ BECTON DICKINSON GMBH 2033567

Patients

Seq Age Sex Outcome Treatment
1 Unknown