BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD
Report
- Report Number
- 9680577-2022-00053
- Event Type
- Malfunction
- Date Received
- April 8, 2022
- Date of Event
- March 16, 2022
- Report Date
- April 28, 2022
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- JSJ
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD. CATALOG NUMBER 221353 WITH 510K NUMBER PREAMENDMENT. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY IT WAS REPORTED THAT 54 PLATES WOULD HAVE SHOW CONTAMINATION. THE COMPLAINT TRENDS WERE REVIEWED AND SIMILAR COMPLAINTS WERE FOUND FOR THIS LOT NUMBER. THEREFORE, A MRB (MATERIAL REVIEW BOARD) WAS PERFORMED AND THIS ISSUE WILL BE INVESTIGATED WITHIN A CAPA (CORRECTIVE AND PREVENTIVE ACTION) 4344556. THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. PICTURES WERE PROVIDED SHOWING THE REPORTED CONTAMINATION. THE RETENTION SAMPLES WERE REVIEWED AND THE PLATES WERE FOUND TO BE WITHOUT DEVIATION. AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. PLEASE NOTE THAT THIS PRODUCT DOES NOT HAVE AN SAL CLAIM. IT IS FILLED ASEPTICALLY, THEREFORE AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. WE ONLY RELEASE PRODUCT BATCHES TO THE MARKET WITH AN AQL (ACCEPTABLE QUALITY LEVEL) = 1.5. ALTHOUGH THIS CONTAMINATION LEVEL IS VERY LOW, WE CANNOT COMPLETELY EXCLUDE THAT A CUSTOMER MAY RECEIVE A CONTAMINATED PLATE. AS SEVERAL COMPLAINTS WERE REPORTED FOR THIS PRODUCT LOT NUMBER REGARDING CONTAMINATION, A MRB (MRB 22-006) AND CAPA (# 4344556) WERE CREATED FOR FURTHER EVALUATION. THE ROOT CAUSE INVESTIGATION IS CURRENTLY ONGOING WITHIN CAPA 4344556. BASED UPON OUR INVESTIGATION THE COMPLAINT WAS CONFIRMED.
IT WAS REPORTED THAT THERE WERE 54 BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD THAT HAD VISIBLE CONTAMINATION ON THE AGAR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE LAST DELIVERY WE RECEIVED FROM YOU OF COLUMBIA CNA AGAR WITH 5% SB, IMPROVED IS LARGELY CONTAMINATED. EVERY 10 PLATES WE HAVE TO DISCARD 3-6 BECAUSE THERE IS SOME FORM OF SIDE CONTAMINATION VISIBLE ON THE AGAR.
IT WAS REPORTED THAT THERE WERE 54 BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD THAT HAD VISIBLE CONTAMINATION ON THE AGAR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE LAST DELIVERY WE RECEIVED FROM YOU OF COLUMBIA CNA AGAR WITH 5% SB, IMPROVED IS LARGELY CONTAMINATED. EVERY 10 PLATES WE HAVE TO DISCARD 3-6 BECAUSE THERE IS SOME FORM OF SIDE CONTAMINATION VISIBLE ON THE AGAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1300425 | BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD | CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL | JSJ | BECTON DICKINSON GMBH | 2033567 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |