FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 14048807 · Received April 7, 2022

Report

Report Number
2032227-2022-176943
Event Type
Malfunction
Date Received
April 7, 2022
Date of Event
April 5, 2022
Report Date
July 11, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000283520
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

RETAINER = MISSING. THE PUMP WAS RETURNED FOR COSMETIC DAMAGE LOCATED ON THE BOTTOM OF THE PUMP, CRACKED RETAINER, AND THE BATTERY CAP MISSING THE METAL CONTACT FOUND ON (B)(6) 2022. THE PUMP WAS RECEIVED WITHOUT THE ORIGINAL BATTERY CAP. THE PUMP WAS RECEIVED WITH MISSING RETAINER, MISSING RESERVOIR TUBE O-RING, AND PARTIALLY BROKEN RESERVOIR TUBE LIP. UNABLE TO PERFORM THE DISPLACEMENT TEST OR LOCK A TEST P-CAP INTO THE RESERVOIR COMPARTMENT DUE TO MISSING RETAINER, MISSING RESERVOIR TUBE O-RING, AND PARTIALLY BROKEN RESERVOIR TUBE LIP. THE FOLLOWING WERE NOTED DURING PHYSICAL INSPECTION: PARTIALLY BROKEN RESERVOIR TUBE LIP, CRACKED END CAP, STAINED KEYPAD OVERLAY, MISSING RETAINER, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, AND MISSING RESERVOIR TUBE O-RING. UNABLE TO VERIFY BATTERY CAP DAMAGE DUE TO MISSING BATTERY CAP. CRACKED END CAP IS CONFIRMED. BATTERY CAP DAMAGE IS UNKNOWN DUE TO MISSING BATTERY CAP. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS DAMAGED. THE RESERVOIR WAS NOT ABLE TO LOCK IN PLACE. CUSTOMER ALSO REPORTED DAMAGE TO THE BATTERY CAP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLE SHOOTING WAS PERFORMED AND THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457518 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG4RY03 000000763000283520

Patients

Seq Age Sex Outcome Treatment
1 30 YR Female