FDA Adverse Event Malfunction Summary report: N

XPS® BLADE

MDR report key: 14047468 · Received April 7, 2022

Report

Report Number
1045254-2022-00178
Event Type
Malfunction
Date Received
April 7, 2022
Date of Event
December 1, 2021
Report Date
May 4, 2022
Manufacturer
MEDTRONIC XOMED INC.
Product Code
EQJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE BUR WAS DAMAGED. SITE HAD TO OPEN TWO MORE TO COMPLETE THE CASE. THERE WAS NO PATIENT IMPACT. THE EVENT OCCURRED DURING A FESS PROCEDURE, INTRA-OP. THERE WAS A DELAY OF LESS THAN 1 HOUR.

Description of Event or Problem · 0

ON FOLLOW-UP, IT WAS REPORTED THAT THE BUR WAS BROKEN WHILE BEING USED. THE BREAK OCCURRED AT THE BEND/JOINT OF THE BUR. BROKEN PARTS WERE CONTAINED IN THE BUR ITSELF THAT PREVENTED THE FRAGMENTS FROM CONTACTING THE PATIENT. THERE WERE NO FRAGMENT IN PATIENT. THERE WERE NO SMALL PIECES THAT CUSTOMER WOULD CONSIDER A FRAGMENT, BUT BUR DID BREAK INTO 2 PIECES. SALES REPRESENTATIVE ATTENDED CASE AFTER HEARING ABOUT THESE CASES, AND THAT CASE WENT FINE, NO ISSUE OR BREAKAGE OF THE BURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
939338 XPS® BLADE BUR, EAR, NOSE AND THROAT EQJ MEDTRONIC XOMED INC. 1882569HS

Patients

Seq Age Sex Outcome Treatment
1 Female