LASER APPLICATOR
Report
- Report Number
- 1022819-2009-00150
- Event Type
- Injury
- Date Received
- June 5, 2009
- Date of Event
- April 1, 2009
- Report Date
- May 12, 2009
- Manufacturer
- CHATTANOOGA GROUP
- Product Code
- ILY
- PMA / PMN Number
- 040662
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A DEVICE EVAL WAS PERFORMED BY OUR ENGINEERING DEPARTMENT. ROOT CAUSE IS UNK BECAUSE THE DEVICE PERFORMED TO SPECIFICATION AND TO ITS INTENDED USE. THE ENGINEERING DEPARTMENT DID NOT FIND ANY PROBLEMS WITH THE DEVICE. CONCLUSION: UNIT PASSED ALL TEST CONDUCTED, NO ANOMALIES WERE FOUND. CHECKING UNIT WITH THE LASER APPLICATOR THE UNIT AND APPLICATOR OUTPUT WAS 1.34W'S. THE UNIT MEETS ALL MANUFACTURES SPECIFICATIONS.
THE PT WAS BEING TREATED FOR MUSCLE SPASMS AND INFLAMMATION TO THE BILATERAL TRAPS WITH LASER WHEN THEY RECEIVED A BURN. THE TREATMENT WAS SET FOR 1.44 MINUTES WITH A DOSAGE OF 7.5 J-CM SQUARED. THERE WAS NO PROTECTIVE BARRIER BETWEEN THE SKIN AND THE APPLICATOR. THE APPLICATOR WAS SLIGHTLY ABOVE THE SKIN BUT COULD HAVE BEEN TOUCHING THE PT'S SKIN DURING THE TREATMENT. THE PT DID COMPLAIN OF PAIN DURING THE TREATMENT. AFTER THE TREATMENT, SEVERAL BLISTERS OCCURRED ON BOTH SIDES OF THE TRAPS. THE PT DID NOT REQUIRE MEDICATION OR MEDICAL ATTENTION FOR THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LASER APPLICATOR | ILY | CHATTANOOGA GROUP | 27808 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR | Other |