FDA Adverse Event Injury Summary report: N

IMPL TWIST MP-1 3.75 MM 1 3 MM

MDR report key: 14039723 · Received April 7, 2022

Report

Report Number
0002023141-2022-00828
Event Type
Injury
Date Received
April 7, 2022
Date of Event
October 28, 2021
Report Date
April 7, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K013494
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PREMARKET IDENTIFICATION K943604. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION 26 WAS REMOVED DUE TO PERI-IMPLANTITIS WITH ABUNDANT BONE LOSS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2341180 IMPL TWIST MP-1 3.75 MM 1 3 MM DENTAL IMPLANT DZE ZIMMER DENTAL 60843512

Patients

Seq Age Sex Outcome Treatment
1 85 YR Female Required Intervention