FDA Adverse Event Malfunction Summary report: N

SUCTION TUBE MCEN78J30 DIA3MM JANKOWSKI

MDR report key: 14031356 · Received April 6, 2022

Report

Report Number
3003249645-2022-00011
Event Type
Malfunction
Date Received
April 6, 2022
Date of Event
March 16, 2022
Report Date
May 16, 2022
Manufacturer
INTEGRA MICROFRANCE S.A.S.
Product Code
JZF
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE JANKOWSKI SUCTION TUBE (MCEN78J30) WAS RETURNED FOR EVALUATION: THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES RELATED TO THE REPORTED FAILURE WERE OBSERVED. FAILURE ANALYSIS: EVALUATION OF THE SUCTION TUBE VERIFIED THE COMPLAINT REPORTED BY THE CUSTOMER AS VALID. THE DEVICE WAS BROKEN NEAR THE LASER WELDING. THE TUBE WAS BENT NEAR THE BREAKAGE AREA. NO MANUFACTURING DEFECT WAS FOUND. ROOT CAUSE ANALYSIS: THIS ISSUE WAS LIKELY DUE TO AN EXCESSIVE EFFORT APPLIED ON THE DEVICE DURING THE USE, THE REPROCESSING OR THE STORAGE OF THE DEVICE.

Description of Event or Problem · 0

THIS REPORT IS 2 OF 2 LINKED TO MFG REPORT NUMBER 3003249645-2022-00010: A FACILITY REPORTED THAT DURING A SINUS INTERVENTION PROCEDURE, AFTER THE SUCTION TUBE (MCEN78J30) WAS INTRODUCED INTO PATIENT'S NOSE, THE CANNULA BROKE WHEN SUCTION WAS INITIATED. THE ISSUE OCCURRED TWICE DURING THE SURGERY, AND A 3RD CANNULA WAS USED. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE CANNULA BROKE AT CURVATURE (BENT) IN THE PATIENT'S NOSE, BUT NO LESION/INJURY FOR THE PATIENT. THE BROKEN PARTS CAME OUT OF THE NOSE WITHOUT ANY INTERVENTION (BY THEMSELVES). THE EVENT DID NOT LEAD TO INCREASED SURGERY TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
812987 SUCTION TUBE MCEN78J30 DIA3MM JANKOWSKI PFM11 JZF INTEGRA MICROFRANCE S.A.S. 5244191

Patients

Seq Age Sex Outcome Treatment
1 Unknown