HAHN TAPERED IMPLANT 4.3 X 8 MM
Report
- Report Number
- 3011649314-2022-00148
- Event Type
- Injury
- Date Received
- April 6, 2022
- Date of Event
- March 10, 2022
- Report Date
- January 8, 2025
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K143353
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE HAS NOT BEEN RETURNED. IF/ WHEN THE DEVICE IS RETURNED AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT'S WEIGHT IS NOT RECORDED AT THE TIME OF VISIT.
THE DEVICE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS THE DHR WAS REVIEWED FOR LOT# 6090528 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS THE PACKAGED STOCK PRODUCT IS NOT APPLICABLE FOR REVIEW SINCE NO DEFECT WAS OBSERVED FROM THE RETURNED PRODUCT. INVESTIGATION METHODS/RESULTS THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE DEVICE WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø4.3 X 8 MM (70-1154-IMP0009) USING RADIOGRAPHIC TEMPLATE ((B)(4)). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. SIGNIFICANT BONE DEBRIS WAS OBSERVED ON THE IMPLANT. ROOT CAUSE "LACK OF PRIMARY STABILITY" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 3027904 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU 3027904 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION."THE IFU ALSO CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN."
D6A AND D6B IN THE INITIAL REPORT WERE INADVERTENTLY REPORTED AS 3-10-2022. THE DEVICE IS NOT CONSIDERED TO HAVE BEEN IMPLANTED AS THE PATIENT DID NOT LEAVE THE STERILE FIELD WITH THE IMPLANT IMPLANTED. CAPA 23-006. MANUFACTURER REFERENCE: (B)(4).
CORRECTED INFORMATION: G1, G3. G3 IN THE PREVIOUS REPORT WAS INADVERTENTLY REPORTED AS 05/22/2024, HOWEVER, THE CORRECT DATE IS 03/10/2022. CAPA CA-00016 MANUFACTURER REFERENCE: (B)(4).
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT LACKED STABILITY. THE PATIENT'S BONE TYPE IS IV, AND THERE IS NO MEDICAL OR DENTAL HISTORY NOTED. THE PATIENT PRESENTED ON (B)(6) 2022 FOR PRIMARY PROCEDURE ON TOOTH #5. DURING PLACEMENT, THE PROVIDER WAS UNABLE TO ACHIEVE STABILITY DUE TO POOR BONE QUALITY, AND THE DEVICE WAS REMOVED. THE IMPLANT WAS REPLACED WITH ONE OF THE SAME SIZE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1440883 | HAHN TAPERED IMPLANT 4.3 X 8 MM | HAHN TAPERED IMPLANT | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1154-IMP0009 | 6090528 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Female | Required Intervention |