FDA Adverse Event Injury Summary report: N

SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY)

MDR report key: 14022190 · Received April 5, 2022

Report

Report Number
MW5108799
Event Type
Injury
Date Received
April 5, 2022
Date of Event
March 1, 2022
Report Date
March 3, 2022
Manufacturer
UNK
Product Code
MEG
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
TX, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THIS WAS NOTED TO BE A PRODUCT ISSUE AND DID NOT REALLY AFFECT PATIENT, WHICH IS WHY A TEST PATIENT WAS SED. NURSE GIVING COVID VACCINE WITH SAFETY SYRINGE "VANISH POINT NEEDLES" SUPPLIED BY PHARMACY DEPARTMENT, NOTICED THAT SAFETY SYRINGES ARE NOT FUNCTIONING CORRECTLY. THE NEEDLE ON THE SYRINGE IS NOT RETRACTING INTO THE SYRINGE FOR SAFETY. NURSING PRACTICE WHEN THIS HAPPENS IS TO PLACE SYRINGE WITH NEEDLE EXPOSED STRAIGHT TO SHARPS CONTAINER, IN FRONT OF VETERAN WITH CAUTION. THIS WAS REPORTED AND RESPECTIVE AFFECTED INDIVIDUALS NOTIFIED, INCLUDING MEDLINE AND PATIENT SAFETY ON 3/9/2022, ALONG WITH ACTION PLANS. PER INFORMATION FROM PHARMACY: THE PHARMACY EMERGENCY PREPAREDNESS MANAGER MET WITH THE PROCUREMENT MANAGER AND VACCINE CLINIC NURSE MANAGER. IT WAS SUSPECTED THAT THE VANISHPOINT SYRINGE LOT NUMBER WHICH HAS BEEN IN USE SINCE (B)(6) (LOT #:0210701) ARE MALFUNCTIONING. ALL OF THESE SYRINGES HAVE BEEN QUARANTINED UNTIL FURTHER NOTICE. A TOTAL OF 1700 NEW VANISHPOINT SYRINGES WITH THREE NEW LOT NUMBERS HAVE BEEN ACQUIRED. MONITORING WILL OCCUR TO ENSURE MALFUNCTIONING IS NOT OCCURRING WITH THESE LOTS. IT IS BELIEVED THAT THE LOT #0210701 HAS CAUSED 15 INCIDENTS SINCE LATE (B)(6) 2022. VANISH POINT SYRINGE/NEEDLE NOT RETRACTING; SUSPECTED LOT# 0210701.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1064738 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
1064739 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
1064740 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
1064741 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940097 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940098 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940099 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940100 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940101 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940102 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940103 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940104 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940105 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940106 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701
940107 SYRINGE (SYRINGE LMUNDL 27G 0.5IN SAFETY) SYRINGE, ANTISTICK MEG UNK G210701

Patients

Seq Age Sex Outcome Treatment
1 40 YR Male