FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED STD STEM SIZE 2

MDR report key: 14019162 · Received April 6, 2022

Report

Report Number
3005180920-2022-00238
Event Type
Injury
Date Received
April 6, 2022
Date of Event
March 9, 2022
Report Date
April 6, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804083
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR. PARTIAL REVISION PERFORMED 7 YEARS AFTER PRIMARY TOTAL HIP ARTHROPLASTY IN A MALE PATIENT WITH A VERY NARROW FEMORAL CANAL AND GOOD CORTICES. FROM THE RADIOGRAPHIC IMAGE, A HETEROTOPIC OSSIFICATION (HO) IS VISIBLE ON THE GREAT THROCANTER, BUT ITS RELATIONSHIP WITH THE ADVERSE EVENT IS DOUBTFUL. THERE ARE SIGNS OF A PREVIOUS DIAPHYSEAL FEMORAL FRACTURE (DATE OF OCCURRENCE UNKNOWN), WHICH MAY HAVE CHANGED OR INFLUENCED THE DIAPHYSEAL MECHANICS AND THEREFORE CONTRIBUTED TO THE POTTING OF THE STEM. THE QUALITY OF THE IMAGES SUPPLIED DOES NOT ALLOW FURTHER CONCLUSIONS TO BE DRAWN. BATCH REVIEW PERFORMED ON 10 MARCH 2022. LOT 146952: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-JAN-2015. EXPIRATION DATE: 2019-11-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ANOTHER SIMILAR REPORTED EVENT SINCE 1-1-2018.

Description of Event or Problem · 0

AT 7 YEARS AND 1 MONTHS POST PRIMARY, THE SURGEON REVISED THE PATIENT FOR STEM LOOSENING. STEM AND HEAD REVISED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2339483 STEM: AMISTEM H HA COATED STD STEM SIZE 2 HIP STEM CEMENTLESS LZO MEDACTA INTERNATIONAL SA 01.18.132 146952 07630030804083

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention