FDA Adverse Event Malfunction Summary report: N

LEUKOSTRIP 13X102MM CTN 50

MDR report key: 14018435 · Received April 5, 2022

Report

Report Number
8043484-2022-00120
Event Type
Malfunction
Date Received
April 5, 2022
Date of Event
March 14, 2022
Report Date
July 12, 2022
Manufacturer
SMITH & NEPHEW MEDICAL LTD.
Product Code
MZR
UDI-DI
05000223449186
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: CASE-(B)(4).

Additional Manufacturer Narrative · 0

H3, H6: THE PRODUCT HAS BEEN RETURNED AND EVALUATED, THE EVENT IS CONFIRMED. VISUAL INSPECTION OF THE DRESSING HAS CONFIRMED THE REPORTED CONTAMINATION, THIS HAS BEEN CONFIRMED TO BE ADDITIONAL ADHESIVE, THIS IS NOT ANY OTHER FORM OF CONTAMINATION. A DOCUMENTATION REVIEW HAS REVEALED PREVIOUSLY CASES WITH A SIMILAR NATURE, THERE ARE NO OPEN CORRECTIVE ACTIONS. THE MANUFACTURING RECORDS CONFIRM THE DEVICE MET THE FINAL PRODUCT SPECIFICATIONS AND CONTAIN NO CONTRIBUTORY FACTORS TO THE REPORTED EVENT, THE ASSOCIATED RISK FILES MITIGATE THIS EVENT TYPE. A REVIEW OF THE MANUFACTURING PROCESS HAS REVEALED THE PROBABLE CAUSE. A PROBABLE CAUSE INCLUDES THE DEFECT WAS NOT PICKED OUT FROM RAW MATERIAL PRIOR TO PRODUCTION, RETRAINING HAS BEEN CONDUCTED, TO REDUCE THE PROBABILITY OF FUTURE OCCURRENCES. NO CORRECTIVE ACTIONS ARE DEEMED NECESSARY, AT THIS STAGE, SMITH AND NEPHEW WILL CONTINUE TO MONITOR FOR ADVERSE TRENDS RELATING TO THE REPORTED EVENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING SET UP FOR WOUND TREATMENT, WAS CONFIRMED A FOREIGN SUBSTANCE ADHERED ON THE STRIP IN ONE DRESSING OF A LEUKOSTRIP 13X102MM CTN 50. AS THIS WAS NOTICED BEFORE TREATMENT, THERE WAS NOT ANY PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2401491 LEUKOSTRIP 13X102MM CTN 50 TOPICAL APPROXIMATION SYSTEM MZR SMITH & NEPHEW MEDICAL LTD. 66002880 2053A 05000223449186

Patients

Seq Age Sex Outcome Treatment
1 Unknown