BD¿ ¿COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD
Report
- Report Number
- 9680577-2022-00046
- Event Type
- Malfunction
- Date Received
- April 5, 2022
- Date of Event
- March 10, 2022
- Report Date
- April 28, 2022
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- JSJ
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY IT WAS REPORTED THAT SOME PLATES WOULD HAVE BEEN CONTAMINATED WITH STAPH. EQUORUM. THE COMPLAINT TRENDS WERE REVIEWED AND SIMILAR COMPLAINTS WERE FOUND FOR THIS LOT NUMBER. THEREFORE, A MRB (MATERIAL REVIEW BOARD) WAS PERFORMED AND THIS ISSUE WILL BE INVESTIGATED WITHIN A CAPA (CORRECTIVE AND PREVENTIVE ACTION) 4344556. THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. PICTURES WERE PROVIDED SHOWING THE REPORTED CONTAMINATION. THE RETENTION SAMPLES WERE REVIEWED AND THE PLATES WERE FOUND TO BE WITHOUT DEVIATION. AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. PLEASE NOTE THAT THIS PRODUCT DOES NOT HAVE AN SAL CLAIM. IT IS FILLED ASEPTICALLY, THEREFORE AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. WE ONLY RELEASE PRODUCT BATCHES TO THE MARKET WITH AN AQL (ACCEPTABLE QUALITY LEVEL) = 1.5. ALTHOUGH THIS CONTAMINATION LEVEL IS VERY LOW, WE CANNOT COMPLETELY EXCLUDE THAT A CUSTOMER MAY RECEIVE A CONTAMINATED PLATE. AS SEVERAL COMPLAINTS WERE REPORTED FOR THIS PRODUCT LOT NUMBER REGARDING CONTAMINATION, A MRB (MRB 22-006) AND CAPA (# 4344556) WERE CREATED FOR FURTHER EVALUATION. THE ROOT CAUSE INVESTIGATION IS CURRENTLY ONGOING WITHIN CAPA 4344556. BASED UPON OUR INVESTIGATION THE COMPLAINT WAS CONFIRMED. H3 OTHER TEXT : SEE H10.
IT HAS BEEN REPORTED THAT THE BD¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD HAS BEEN FOUND CONTAMINATED. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: PLATES CONTAMINATED WITH STAPHYLOCOCCUS EQUOROM.
IT HAS BEEN REPORTED THAT THE BD¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD HAS BEEN FOUND CONTAMINATED. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: PLATES CONTAMINATED WITH STAPHYLOCOCCUS EQUOROM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2049046 | BD¿ ¿COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD | BD¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD | JSJ | BECTON DICKINSON GMBH | 2033567 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |