FDA Adverse Event Injury Summary report: N

FONAR INDOMITABLE AKA UPRIGHT MRI

MDR report key: 1401591 · Received June 12, 2009

Report

Report Number
2432211-2009-00001
Event Type
Injury
Date Received
June 12, 2009
Date of Event
May 15, 2009
Report Date
June 11, 2009
Manufacturer
FONAR CORPORATION
Product Code
LNH
PMA / PMN Number
K002490
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE BURN SITES WERE IDENTICAL TO ELECTRICAL CONTACTS ON THE PATIENT'S BED. IN SPITE OF CAUTION LABELS ON THE BED AND INSTRUCTIONS IN THE USER MANUAL INDICATING CONTACT WITH THE SKIN MAY INDUCE RF BURNS DURING SCANNING AND REQUIRING THE CONTACTS BE COVERED BY THE PROVIDED SAFETY COVERS OR BED CUSHION, THE PATIENT WAS POSITIONED SO THAT BARE SKIN WAS IN DIRECT CONTACT WITH THE ELECTRICAL CONTACTS. NEITHER THE SAFETY COVERS NOR BED CUSHION WERE USED TO COVER THE ELECTRICAL CONTACTS. WE FEEL THAT THIS IS A CASE OF USER ERROR, AND NOT A DEFECT IN THE DEVICE. IF THE SYSTEM IS OPERATED IN THE CORRECT MANNER, THIS TYPE OF INJURY SHOULD NOT OCCUR TO ANY PATIENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT HAD BEEN POSITIONED IN AN UPRIGHT MRI FOR A LUMBAR AND SACRUM EXAM. THE PATIENT CHANGED INTO A PATIENT GOWN WHICH WAS OPEN AT THE BACK AND ENDED AT THE KNEE. THE PATIENT WAS POSITIONED ON THE BED SLAB ON HER BACK WITH A BED CUSHION POSITIONED UNDERNEATH HER. THE CUSHION ENDED AT THE PATIENT'S WAIST. THE PATIENT WAS SITTING ON THE PATIENT SEAT AND THE BED WAS UPRIGHT FOR THE LUMBAR SCAN. THE PATIENT WAS STANDING ON THE PATIENT SEAT IN THE RECUMBENT POSITION, FACE UP, FOR THE SACRUM SCAN. BOTH SCANS WERE COMPLETED WITH NO COMPLICATIONS OR COMMENTS FROM THE PATIENT. UPON LEAVING THE MRI SUITE, THE PATIENT MADE THE COMMENT THAT THE MACHINE BURNED HER. THE TECHNOLOGIST EXAMINED THE PATIENT AND FOUND A RED BURN APPROXIMATELY 5-6 INCHES IN DIAMETER ON THE RIGHT LATERAL CALF, 1-1.5 INCHES BELOW HER KNEE. THE PATIENT SOUGHT MEDICAL ATTENTION FROM HER PRIMARY CARE PHYSICIAN AND WAS FOUND TO HAVE 2 BURNS, ONE OF WHICH WAS A 3RD DEGREE BURN 3/16 INCH IN DIAMETER. THE PHYSICIAN TREATED HER WITH A TOPICAL OINTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FONAR INDOMITABLE AKA UPRIGHT MRI MAGNETIC RESONANCE IMAGING SCANNER LNH FONAR CORPORATION UPRIGHT MRI

Patients

Seq Age Sex Outcome Treatment
1 43 YR Other