FONAR INDOMITABLE AKA UPRIGHT MRI
Report
- Report Number
- 2432211-2009-00001
- Event Type
- Injury
- Date Received
- June 12, 2009
- Date of Event
- May 15, 2009
- Report Date
- June 11, 2009
- Manufacturer
- FONAR CORPORATION
- Product Code
- LNH
- PMA / PMN Number
- K002490
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE BURN SITES WERE IDENTICAL TO ELECTRICAL CONTACTS ON THE PATIENT'S BED. IN SPITE OF CAUTION LABELS ON THE BED AND INSTRUCTIONS IN THE USER MANUAL INDICATING CONTACT WITH THE SKIN MAY INDUCE RF BURNS DURING SCANNING AND REQUIRING THE CONTACTS BE COVERED BY THE PROVIDED SAFETY COVERS OR BED CUSHION, THE PATIENT WAS POSITIONED SO THAT BARE SKIN WAS IN DIRECT CONTACT WITH THE ELECTRICAL CONTACTS. NEITHER THE SAFETY COVERS NOR BED CUSHION WERE USED TO COVER THE ELECTRICAL CONTACTS. WE FEEL THAT THIS IS A CASE OF USER ERROR, AND NOT A DEFECT IN THE DEVICE. IF THE SYSTEM IS OPERATED IN THE CORRECT MANNER, THIS TYPE OF INJURY SHOULD NOT OCCUR TO ANY PATIENT.
IT WAS REPORTED THAT A PATIENT HAD BEEN POSITIONED IN AN UPRIGHT MRI FOR A LUMBAR AND SACRUM EXAM. THE PATIENT CHANGED INTO A PATIENT GOWN WHICH WAS OPEN AT THE BACK AND ENDED AT THE KNEE. THE PATIENT WAS POSITIONED ON THE BED SLAB ON HER BACK WITH A BED CUSHION POSITIONED UNDERNEATH HER. THE CUSHION ENDED AT THE PATIENT'S WAIST. THE PATIENT WAS SITTING ON THE PATIENT SEAT AND THE BED WAS UPRIGHT FOR THE LUMBAR SCAN. THE PATIENT WAS STANDING ON THE PATIENT SEAT IN THE RECUMBENT POSITION, FACE UP, FOR THE SACRUM SCAN. BOTH SCANS WERE COMPLETED WITH NO COMPLICATIONS OR COMMENTS FROM THE PATIENT. UPON LEAVING THE MRI SUITE, THE PATIENT MADE THE COMMENT THAT THE MACHINE BURNED HER. THE TECHNOLOGIST EXAMINED THE PATIENT AND FOUND A RED BURN APPROXIMATELY 5-6 INCHES IN DIAMETER ON THE RIGHT LATERAL CALF, 1-1.5 INCHES BELOW HER KNEE. THE PATIENT SOUGHT MEDICAL ATTENTION FROM HER PRIMARY CARE PHYSICIAN AND WAS FOUND TO HAVE 2 BURNS, ONE OF WHICH WAS A 3RD DEGREE BURN 3/16 INCH IN DIAMETER. THE PHYSICIAN TREATED HER WITH A TOPICAL OINTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FONAR INDOMITABLE AKA UPRIGHT MRI | MAGNETIC RESONANCE IMAGING SCANNER | LNH | FONAR CORPORATION | UPRIGHT MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Other |