FACTOR V LEIDEN KIT
Report
- Report Number
- 2243471-2009-00013
- Date Received
- June 12, 2009
- Date of Event
- April 27, 2009
- Report Date
- May 20, 2009
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC
- Product Code
- NPQ
- PMA / PMN Number
- K033607
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
REVIEW OF THE PRODUCT LABELING IDENTIFIED THE FOLLOWING TEXT WITHIN THE QUALITY CONTROL SECTION OF THE PACKAGE INSERT: "THE ASSAY RESULT FOR THE FACTOR V LEIDEN CT SHOULD ALWAYS BE 'HET'. IF THIS IS NOT OBSERVED, THE RUN IS INVALID AND PCR MUST BE REPEATED." THERE IS NO MEDICAL RISK ASSOCIATED WITH A FAILED CONTROL AS THE GENERATED RESULTS ARE INVALID AND NOT TO BE REPORTED. THE PRIMARY MEDICAL TREATMENT DECISION THAT WOULD BE BASED ON THE RESULT OF THE LC FACTOR V LEIDEN TEST WOULD BE WHETHER TO DISCONTINUE ANTICOAGULATION AFTER A STANDARD 3 - 6 MONTHS OR TO CONTINUE ANTICOAGULATION INDEFINITELY. SINCE THE PATIENTS WOULD ALREADY BE UNDER TREATMENT WHEN THE TEST WAS BEING PERFORMED, THERE IS NO RISK OF DELAYING TREATMENT DUE TO A FACTOR V CONTROL FAILURE. THE TREATING CLINICIAN WOULD CONTINUE THE ANTICOAGULATION THERAPY UNTIL A VALID FACTOR V ASSAY RESULT WAS REPORTED. INVESTIGATION INTO THE ROOT CAUSE OF THE CONTROL FAILURES WAS PERFORMED. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO THE GENERATION OF SUPERCOILING OF THE WILDTYPE AND MUTATION PLASMIDS IN THE CONTROL THAT CAUSE UNEQUAL AMPLIFICATION RESULTS BETWEEN THE TWO PLASMIDS, AND THEREFORE, THE RESULT IS INTERPRETED AS INVALID. DUE TO THE LOW LEVEL OF RISK ASSOCIATED WITH A FAILED CONTROL, FURTHER MITIGATION IS NOT BEING CONSIDERED. (B)(4).
THE CUSTOMER REPORTED 6 RUN FAILURES, WHICH AFFECTED 13 PATIENT RESULTS (NOT SPECIFIED WHETHER THESE WERE 13 UNIQUE PATIENTS), DUE TO AN INVALID FACTOR V LEIDEN CONTROL TEMPLATE (FVL CT). THE CUSTOMER WAS PERFORMING AN OFF-LABEL PRACTICE BY REPORTING TEST RESULTS FROM TEST RUNS WITH INVALID CONTROLS; THESE TEST RUNS SHOULD HAVE BEEN INVALIDATED AS PER THE FACTOR V LEIDEN KIT LABELING. REPORTING INVALID TEST RESULTS MAY RESULT IN THE RELEASE OF ERRONEOUS, BUT BELIEVABLE TEST RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FACTOR V LEIDEN KIT | FACTOR V LEIDEN DNA MUTATION DETECTION SYSTEMS, PRODUCT CODE: NPQ | NPQ | ROCHE MOLECULAR SYSTEMS, INC | 14674120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |