FDA Adverse Event Malfunction Summary report: N

INSERTION GUIDE

MDR report key: 13999739 · Received April 4, 2022

Report

Report Number
2939274-2022-01185
Event Type
Malfunction
Date Received
April 4, 2022
Report Date
April 4, 2022
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
FZH
UDI-DI
10886982190697
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT#: (B)(4). IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. REPORTER IS A J&J SALES REPRESENTATIVE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCE'S WERE IDENTIFIED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: H3, H6 INVESTIGATION SUMMARY SERVICE AND REPAIR EVALUATION: THE CUSTOMER REPORTED THE INSERTION GUIDE BROKE. THE REPAIR TECHNICIAN REPORTED MISSING SCREW. MISSING PARTS IS THE REASON FOR REPAIR. THE FOLLOWING PARTS WERE REPLACED: SCREW. THE ITEM WAS REPAIRED PER THE INSPECTION SHEET, PASSED SYNTHES FINAL INSPECTION ON 06-APR-2022 AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. ATTACHED SERVICE RECORD ROUTER COMPLETED THROUGH OPERATION 30. FINALIZED SERVICE RECORD WILL BE ARCHIVED IN TUNGSTEN DOCUMENT MANAGEMENT SYSTEM. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. H4 DEVICE HISTORY PART # 323.050 SYNTHES LOT # 1295110 SUPPLIER LOT # 1295110 RELEASE TO WAREHOUSE DATE: 05 OCT 2004 MANUFACTURING LOCATION: MONUMENT NO NCR'S WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE INSERTION GUIDE BROKE. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE INSERTION GUIDE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1425204 INSERTION GUIDE GUIDE, SURGICAL, INSTRUMENT FZH WRIGHTS LANE SYNTHES USA PRODUCTS LLC 323.050 1295110 10886982190697

Patients

Seq Age Sex Outcome Treatment
1 Unknown