FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 1399932 · Received March 29, 2009

Report

Report Number
2250051-2009-00175
Event Type
Malfunction
Date Received
March 29, 2009
Date of Event
May 27, 2009
Report Date
May 29, 2009
Manufacturer
HAMILTON BONADUZ AG
Product Code
JTC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND CLEANED THE PLUNGER CLAMP AND REPLACED THE TIP CLAMP IN POSITION D4. THE INSTRUMENT WAS TESTED WITHOUT FURTHER PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE REAGENT AND SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTER JTC HAMILTON BONADUZ AG 7003009

Patients

Seq Age Sex Outcome Treatment
1