BD PLASTIPAK¿ SYRINGES
Report
- Report Number
- 3003152976-2022-00133
- Event Type
- Malfunction
- Date Received
- April 4, 2022
- Date of Event
- February 14, 2022
- Report Date
- May 5, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. LOT NUMBER WAS NOT REPORTED; HOWEVER, A POTENTIAL LOT NUMBER WAS PROVIDED. THE INFORMATION FOR THAT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 2111091. MEDICAL DEVICE EXPIRATION DATE: 31OCT2026. DEVICE MANUFACTURE DATE: 10NOV2021. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE SUSPECTED LOT 2111091, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 2111091 WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, THE STOPPERS WERE PROPERLY ASSEMBLED, AND NO LEAKS WERE IDENTIFIED DURING TESTING. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED BEHIND THE STOPPER. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE OF ANTICANCER AGENT THAT PASSES UNDER THE PLUNGER OF THE SYRINGE DURING THE PREPARATION OF ANTICANCER CHEMOTHERAPY.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED BEHIND THE STOPPER. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE OF ANTICANCER AGENT THAT PASSES UNDER THE PLUNGER OF THE SYRINGE DURING THE PREPARATION OF ANTICANCER CHEMOTHERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 81974 | BD PLASTIPAK¿ SYRINGES | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |