BINAXNOW COVID-19 AG CARD
Report
- Report Number
- 1221359-2022-01831
- Event Type
- Malfunction
- Date Received
- April 1, 2022
- Date of Event
- February 10, 2022
- Report Date
- April 1, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 10811877011290
- PMA / PMN Number
- EUA202537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
INVESTIGATION REPORT: TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 170179 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-000 / LOT 170179, CARD DEVICE PART NUMBER 195-430 / LOT 168292. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 170179 SHOWED THAT THE COMPLAINT RATE IS 0.0057 %. ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. RELATED MFR REPORT NUMBERS: 1221359-2022-01762 THROUGH 1221359-2022-01835
CUSTOMER REPORTED SEVENTY-FOUR (74) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT SEVENTY (70) OF SEVENTY-FOUR (74) AND LOT 170179 (TOTAL QUANTITY 16). PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2022 AT 11:31 O'CLOCK WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR NASAL SAMPLE WAS COLLECTED THE SAME DAY, PLACED IN VTM, AND TRANSPORTED VIA CURRIER TO A SEPARATE BRANCH LAB FOR TESTING (PLATFORM: (B)(6)) WHICH GENERATED A NEGATIVE RESULT. THE PATIENT WAS ASYMPTOMATIC. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT. CUSTOMER IS PERFORMING TESTS IN COMMUNITY ORGANIZATIONS (NON-K-12 SCHOOLS, NON LONG-TERM CARE FACILITIES).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1839233 | BINAXNOW COVID-19 AG CARD | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 170179 | 10811877011290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Female |