Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 172101 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-000 / LOT 172101, CARD DEVICE PART NUMBER 195-430WL / LOT 169537. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 172101 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. RELATED MFR REPORT NUMBERS: 1221359-2022-01762 THROUGH 1221359-2022-01835.

Description of Event or Problem · 0

CUSTOMER REPORTED SEVENTY-FOUR (74) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT TWENTY-FOUR (24) OF SEVENTY-FOUR (74) AND LOT 172101 (TOTAL QUANTITY (B)(4)). PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2022 AT 11:52 O'CLOCK WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR NASAL SAMPLE WAS COLLECTED THE SAME DAY, PLACED IN VTM, AND TRANSPORTED VIA CURRIER TO A SEPARATE BRANCH LAB FOR TESTING (PLATFORM: PERKINELMER) WHICH GENERATED A NEGATIVE RESULT. THE PATIENT WAS SYMPTOMATIC. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT. CUSTOMER IS PERFORMING TESTS IN COMMUNITY ORGANIZATIONS (NON-K-12 SCHOOLS, NON LONG-TERM CARE FACILITIES).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1967969 BINAXNOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 172101 10811877011290

Patients

Seq Age Sex Outcome Treatment