FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG CARD

MDR report key: 13982146 · Received April 1, 2022

Report

Report Number
1221359-2022-01826
Event Type
Malfunction
Date Received
April 1, 2022
Date of Event
February 3, 2022
Report Date
April 1, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
10811877011290
PMA / PMN Number
EUA202537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS (B)(4) INC. ON RETAINED KIT LOT 170179 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-000 / LOT 170179, CARD DEVICE PART NUMBER 195-430 / LOT 168292. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 170179 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. RELATED MFR REPORT NUMBERS: 1221359-2022-01762 THROUGH 1221359-2022-01835.

Description of Event or Problem · 0

CUSTOMER REPORTED SEVENTY-FOUR (74) FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD ON MULTIPLE DATES ON KITTED NASAL SWABS. THIS REPORT IS FOR PATIENT SIXTY-FIVE (65) OF SEVENTY-FOUR (74) AND LOT 170179 (TOTAL QUANTITY 16). PATIENT WAS TESTED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2022 AT 10:16 O'CLOCK WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR TESTING (PLATFORM: (B)(6)) WAS PERFORMED THE SAME DAY WHICH GENERATED NEGATIVE RESULTS (SAMPLE TYPE UNKNOWN). THE PATIENT WAS SYMPTOMATIC. NO SERIOUS INJURY, MEDICAL INTERVENTION OR TREATMENT DECISIONS OCCURRED AS A RESULT. CUSTOMER IS PERFORMING TESTS IN COMMUNITY ORGANIZATIONS (NON-K-12 SCHOOLS, NON LONG-TERM CARE FACILITIES).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1228063 BINAXNOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 170179 10811877011290

Patients

Seq Age Sex Outcome Treatment
1 50 YR Male