FDA Adverse Event Injury Summary report: N

OMNILINK ELITE PERIPHERAL STENT SYSTEM

MDR report key: 13978441 · Received April 1, 2022

Report

Report Number
2024168-2022-03471
Event Type
Injury
Date Received
April 1, 2022
Date of Event
March 9, 2022
Report Date
July 15, 2022
Manufacturer
ABBOTT VASCULAR
Product Code
NIO
UDI-DI
08717648178559
PMA / PMN Number
P110043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE ADDITIONAL OMNILINK STENT DELIVERY SYSTEM (SDS) (1012632-29, 1030941) IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE INVESTIGATION DETERMINED THE REPORTED DIFFICULTIES AND SUBSEQUENT REMOVAL OF FOREIGN BODY APPEAR TO BE RELATED TO THE OPERATIONAL CONTEXT OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN, OR LABELING.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE RIGHT COMMON ILIAC ARTERY. DURING ADVANCING OF THE FIRST OMNILINK STENT DELIVERY SYSTEM (SDS) (1012632-39, 1050441), THE STENT DISLODGED, AND A SNARE DEVICE WAS USED TO RETRIEVE THE DISLODGED STENT SUCCESSFULLY. A SECOND OMNILINK SDS (1012632-29, 1030941) WAS ADVANCED INTO THE VESSEL AND THE STENT DISLODGED AS WELL. THE DISLODGED STENT WAS ABLE TO BE RETRIEVED WITH THE DELIVERY SYSTEM BUT A CUT DOWN OF THE VESSEL WAS PERFORMED TO REMOVE THE SECOND DISLODGED STENT. THERE WAS SOME RESISTANCE NOTED DURING ADVANCEMENT DUE TO A SHARP ANGLE IN THE VESSEL. A THIRD OMNILINK SDS WAS USED SUCCESSFULLY TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1013242 OMNILINK ELITE PERIPHERAL STENT SYSTEM PERIPHERAL STENT DELIVERY SYSTEM NIO ABBOTT VASCULAR 1012632-39 1050441 08717648178559

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention