FDA Adverse Event Malfunction Summary report: N

EXACTECH

MDR report key: 13975385 · Received April 1, 2022

Report

Report Number
1038671-2022-00341
Event Type
Malfunction
Date Received
April 1, 2022
Date of Event
September 7, 2018
Report Date
April 1, 2022
Manufacturer
EXACTECH, INC.
Product Code
LHX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AS A RESULT OF AN FDA INSPECTION CONDUCTED IN JAN 2020, EXACTECH (B)(4) HAS COMMITTED TO REMEDIATING 3 YEARS OF COMPLAINTS (2017-2020). THIS MDR IS BEING SUBMITTED AS PART OF THAT REMEDIATION. INFORMATION WAS NOT PROVIDED FOR: PATIENT. ENGINEERING PRINTS OF 02-029-122-2000 WERE REVIEWED. THE OVERLAP BETWEEN THE SNAP RING¿S OD AND THE ID OF THE HOUSING LIP IS .008 FOR THE WORST-CASE SCENARIO. ALTHOUGH THE OVERLAP IS POSITIVE, THE SMALL AMOUNT COULD CAUSE DISENGAGEMENT ROOT CAUSE PER CAPA (B)(4), THE ROOT CAUSE FOR THE BUTTON DISENGAGEMENT WAS LIKELY THE RESULT OF INSUFFICIENT OVERLAP BETWEEN THE SNAP RING AND THE LIP OF THE HOUSING TO PREVENT THE OCCURRENCE OF DISENGAGEMENT. CORRECTIVE ACTION TAKEN: "PER (B)(4), THE COMMITTEE HAS DETERMINED THAT THE DESIGN WILL BE UPDATED, TESTED, AND IMPLEMENTED, AFTER WHICH A STAGED MARKET WITHDRAWAL WAS CONDUCTED. PER CAPA (B)(4): A STOCK AUDIT WAS PERFORMED FOR THE DEVICES LOCATED IN 01 (COC). SEE NPR-40446 AND NPR-45222 FOR DETAILS. DEFECTIVE DEVICES WERE DISPOSITIONED FOR REWORK. THE DESIGN WAS UPDATED TO INCREASE THE OVERLAP OF THE SNAP RING AND THE HOUSING LIP COMPONENTS. SEE DCRTC-18-0401 AND DCRTC-19-0446 FOR DETAILS. THE DEFECTED DEVICES RETURNED FROM THE FIELD ARE TO BE REWORKED. SEE STAGED MARKET WITHDRAWAL OF TRULIANT TIBIAL TAMP GUIDE DOCUMENTED IN THE CAPA FOR DETAILS." IFU 700-096-181: INSTRUMENT INSPECTION: VISUALLY INSPECT THE INSTRUMENTS FOR DAMAGE SUCH AS FRACTURES; CRACKS; GOUGES; DEFORMATION; BURRS; DISCOLORATION, CORROSION, OR RUST; EXCESSIVE COMPONENT WEAR; NICKS ON CUTTING SURFACES, MISSING OR LOOSE COMPONENTS; BLOCKAGES IN CANNULAE, CLEANING HOLES OR OTHER CAVITIES THAT CANNOT BE REMOVED VIA STANDARD CLEANING; WORN OR DIFFICULT TO READ MARKINGS/ENGRAVINGS; OR OTHER APPARENT DAMAGE. CHECK THE FUNCTION OF MECHANISMS BY ACTUATING ANY LEVERS, KNOBS, SWITCHES, CONNECTORS, SLIDING FEATURES, HINGES, OR OTHER MECHANICAL INTERFACE FEATURES. ENSURE SMOOTH OPERATION OF THESE FEATURES OVER THEIR FUNCTIONAL RANGE OF MOTION. IF DAMAGE, WEAR, OR NON-FUNCTIONING/POORLY FUNCTIONING MECHANISMS ARE FOUND, DO NOT USE THE INSTRUMENT, AND CONTACT THE SALES REPRESENTATIVE OR CUSTOMER SERVICE FOR DISPOSITION. IFU STATES: THE SURGEON SHALL BECOME THOROUGHLY FAMILIAR WITH THE TECHNIQUE OF IMPLANTATION OF THE PROSTHESES BY: (1) APPROPRIATE READING OF THE LITERATURE, AND (2) TRAINING IN THE OPERATIVE SKILLS AND TECHNIQUES REQUIRED FOR SURGERY, AND (3) REVIEWING INFORMATION REGARDING USE OF INSTRUMENTATION. IT IS A CLINICAL STANDARD OF PRACTICE IN THE OPERATING ROOM THAT ALL INSTRUMENTS SHOULD BE VISUALLY AND FUNCTIONALLY INSPECTED BEFORE USE, THESE GUIDELINES ARE FROM STANDARD ASSOCIATION OF PERI-OPERATIVE REGISTERED NURSES (AORN) GUIDELINES. SURGEONS ARE TO BE FAMILIAR AND KNOWLEDGEABLE WITH ALL INSTRUMENTATION, DEVICES AND PROFICIENT WITH SURGICAL TECHNIQUES. THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO REASONABLY SUGGEST THE REPORTED EVENT IS RELATED TO ANY MANUFACTURING ISSUES. NO PATIENT ADVERSE EVENT. PER CAPA (B)(4), THE ROOT CAUSE FOR THE BUTTON DISENGAGEMENT WAS LIKELY THE RESULT OF INSUFFICIENT OVERLAP BETWEEN THE SNAP RING AND THE LIP OF THE HOUSING TO PREVENT THE OCCURRENCE OF DISENGAGEMENT.

Description of Event or Problem · 0

IT WAS REPORTED FROM THE US THAT AFTER AN ORTHOPEDIC SURGERY THE TIBIAL TAMP GUIDE WAS FOUND IN TWO PIECES. IT WAS STATED THAT THE HANDLE APPEARED TO WORK PROPERLY DURING THE CASE. WHEN THE AGENT WENT TO PUT THE TRAYS BACK TOGETHER, THE HANDLE WAS IN TWO PIECES. NO PATIENT INVOLVEMENT AT THE TIME OF DISCOVERY. NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1186056 EXACTECH TRULIANT TIB TAMP GUIDE LHX EXACTECH, INC. 105607006

Patients

Seq Age Sex Outcome Treatment
1 Unknown