MINICAP TRANSFER SET
Report
- Report Number
- 1416980-2022-01483
- Event Type
- Malfunction
- Date Received
- March 31, 2022
- Date of Event
- March 1, 2022
- Report Date
- July 14, 2022
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- KDJ
- UDI-DI
- 00085412007731
- PMA / PMN Number
- K152675
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
INITIAL REPORTER FACILITY NAME : (B)(6) HOSPITAL . SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
H10: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, A PHOTOGRAPH OF THE SAMPLE WAS PROVIDED FOR EVALUATION. THE PHOTOGRAPH WAS VISUALLY INSPECTED AND A CRACKED MAIN BODY WAS OBSERVED. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CRACKED MAIN BODY COULD NOT BE DETERMINED; THE DAMAGE WAS CONSISTENT WITH DAMAGE CAUSED BY EXPOSURE TO CHEMICAL AGENTS SUCH AS FOREIGN SOLVENTS, DYES, OR CLEANING AGENTS THAT DAMAGE THE TRANSFER SET MATERIALS. A NONCONFORMANCE HAS BEEN OPENED TO ADDRESS THIS ISSUE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
H10: ONE ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WITH THE NAKED EYE NOTED A DAMAGED/CRACKED MAIN BODY OF THE TWIST CLAMP AND BROKEN OCCLUDER LEGS. THE SAMPLE FAILED TO MEET SPECIFICATION. THE REPORTED CONDITION WAS VERIFIED. THE DAMAGE CRACKED/BROKEN SET IS CONSISTENT WITH DAMAGE CAUSED BY EXPOSURE TO CHEMICAL AGENTS SUCH AS FOREIGN SOLVENTS, DYES, OR CLEANING AGENTS THAT DAMAGE THE TRANSFER SET MATERIALS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS A CRACK IN THE LIGHT BLUE MAIN BODY OF THE TWIST CLAMP ON A MINICAP EXTENDED LIFE PD TRANSFER SET. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 782884 | MINICAP TRANSFER SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE CORPORATION | NA | H21E04030 | 00085412007731 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |