FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 13970634 · Received March 31, 2022

Report

Report Number
1416980-2022-01485
Event Type
Malfunction
Date Received
March 31, 2022
Date of Event
March 8, 2022
Report Date
June 28, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
KDJ
UDI-DI
00085412007731
PMA / PMN Number
K152675
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

H10: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, A PHOTOGRAPH OF THE SAMPLE WAS PROVIDED FOR EVALUATION. THE PHOTOGRAPH WAS VISUALLY INSPECTED AND A CRACKED MAIN BODY WAS OBSERVED. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CRACKED MAIN BODY COULD NOT BE DETERMINED; THE DAMAGE WAS CONSISTENT WITH DAMAGE CAUSED BY EXPOSURE TO CHEMICAL AGENTS SUCH AS FOREIGN SOLVENTS, DYES, OR CLEANING AGENTS THAT DAMAGE THE TRANSFER SET MATERIALS. A NONCONFORMANCE HAS BEEN OPENED TO ADDRESS THIS ISSUE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

H10: THE DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED, AND IT WAS NOTED THAT THE LIGHT BLUE MAIN BODY WAS CRACKED. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED; HOWEVER, THE DAMAGED SET IS CONSISTENT WITH DAMAGE CAUSED BY EXPOSURE TO CHEMICAL AGENTS SUCH AS FOREIGN SOLVENTS, DYES, OR CLEANING AGENTS THAT DAMAGE THE TRANSFER SET MATERIALS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE LIGHT BLUE MAIN BODY OF A MINICAP TRANSFER SET WAS CRACKED. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
782873 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE CORPORATION NA H21E20119 00085412007731

Patients

Seq Age Sex Outcome Treatment
1 Unknown