FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ SYRINGES

MDR report key: 13968510 · Received March 31, 2022

Report

Report Number
3003152976-2022-00126
Event Type
Malfunction
Date Received
March 31, 2022
Date of Event
March 7, 2022
Report Date
May 4, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES, D9: RETURNED TO MANUFACTURER ON: 18-APR-2022. H6: INVESTIGATION SUMMARY: SEVERAL SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, THE BARREL WAS OBSERVED TO BE DAMAGED, VERIFYING THE REPORTED ISSUE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2110020, FINDING ONE ANNOTATION RELATED TO THE ALLEGED DEFECT. DURING THE MARKING PROCESS, IT WAS IDENTIFIED THE MARKING MACHINE WAS PERIODICALLY CAUSING DAMAGE TO THE PRODUCT AS OBSERVED IN THE SAMPLE PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED. THIS OCCURRED 6 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 10 ML SYRINGES ARE LEAKING FROM BUNG.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED. THIS OCCURRED 6 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 10 ML SYRINGES ARE LEAKING FROM BUNG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2189291 BD PLASTIPAK¿ SYRINGES PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2110020

Patients

Seq Age Sex Outcome Treatment
1 Unknown