BD PLASTIPAK¿ SYRINGES
Report
- Report Number
- 3003152976-2022-00126
- Event Type
- Malfunction
- Date Received
- March 31, 2022
- Date of Event
- March 7, 2022
- Report Date
- May 4, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES, D9: RETURNED TO MANUFACTURER ON: 18-APR-2022. H6: INVESTIGATION SUMMARY: SEVERAL SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, THE BARREL WAS OBSERVED TO BE DAMAGED, VERIFYING THE REPORTED ISSUE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2110020, FINDING ONE ANNOTATION RELATED TO THE ALLEGED DEFECT. DURING THE MARKING PROCESS, IT WAS IDENTIFIED THE MARKING MACHINE WAS PERIODICALLY CAUSING DAMAGE TO THE PRODUCT AS OBSERVED IN THE SAMPLE PROVIDED.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED. THIS OCCURRED 6 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 10 ML SYRINGES ARE LEAKING FROM BUNG.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED. THIS OCCURRED 6 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 10 ML SYRINGES ARE LEAKING FROM BUNG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2189291 | BD PLASTIPAK¿ SYRINGES | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2110020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |