FDA Adverse Event Malfunction Summary report: N

ENDOPATH ENDOSCOPIC MULTIFEED STAPLER

MDR report key: 139597 · Received December 19, 1997

Report

Report Number
1527736-1997-03525
Event Type
Malfunction
Date Received
December 19, 1997
Date of Event
November 19, 1997
Report Date
November 21, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE RETURNED WITH NO LOT IDENTIFICATION. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: HEAD ROTATES, ABCDEF)YES; NOSE SHROUD CRACKED/BROKEN, ADEF)NO BC)YES; STAPLES IN NOSE, ADE)NO BCF)YES; STAPLES IN THE TRACK, ABE)NO F)8 AND TRIGGER ENGAGED WITH PRECOCK, ABCDEF)YES. FUNCTIONAL TESTS & RESULTS: CYCLED INSTRUMENT, ABCDE)NO F)YES. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED VISUAL EXAMINATION AND FUNCTIONAL TEST, IT WAS CONCLUDED THAT THE REPORTED "STAPLER FIRED WELL INITIALLY THEN FAILED TO FEED STAPLES" DURING SURGERY OCCURRED DUE TO A SEPARATED NOSE WELD IN INSTRUMENTS (B) AND (C) AND TO AN INCLINED STAPLE IN THE NOSE OF INSTRUMENT (B) AND A TWISTED STAPLE IN THE NOSE OF INSTRUMENT (B). NO CONCLUSION COULD BE REACHED AS TO IF THE STAPLES JAMMED IN THE NOSE OF THE DEVICES CAUSED THE NOSE WELD TO YIELD OR IF THE YIELDED NOSE WELDS CAUSED THE STAPLES TO BECOME JAMMED IN THE NOSE. INSTRUMENTS (A), (D), AND (E) WERE RECEIVED EMPTY IN GOOD CONDITION. INSTURMENT (F) WAS CYCLED AND FIRED THE REMAINING 9 STAPLES WELL FORMED INTERMITTENTLY DUE TO INTERFERENCE WITH THE DRIED BODY FLUIDS IN THE CARTRIDGE NOSE.

Description of Event or Problem · 1

IT WAS REPORTED DURING A LAPAROSCOPIC HERNIA REPAIR THE EMS STAPLER FIRED WELL INITIALLY THEN FAILED TO FEED STAPLES THEREAFTER (X6). 11/21/97 REP REPORTS SIX EMS STAPLERS WERE USED AND WITH EACH STAPLER THE FIRST STAPLE FIRED WITHOUT DIFFICULTY AND THEN JAMMED. A SEVENTH, NEW EMS WAS OPENED TO COMPLETE THE CASE. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ENDOSCOPIC MULTIFEED STAPLER ENDOSCOPIC STAPLER GAG ETHICON ENDO-SURGERY, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other