ENDOPATH ENDOSCOPIC MULTIFEED STAPLER
Report
- Report Number
- 1527736-1997-03523
- Event Type
- Malfunction
- Date Received
- December 19, 1997
- Date of Event
- November 19, 1997
- Report Date
- November 21, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
DEVICE RETURNED WITH NO LOT ID. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: HEAD ROTATES, ABCDEF)YES; NOSE SHROUD CRACKED/BROKEN, ADEF)NO BC)YES; STAPLES IN NOSE, ADE)NO BCF)YES; STAPLES IN THE TRACK, ABE)NO F)8 AND TRIGGER ENGAGED WITH PRECOCK, ABCDEF)YES. FUNCTIONAL TESTS & RESULTS: CYCLED INSTRUMENT, ABCDE)NO F)YES. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO REC'D VISUAL EXAMINATION AND FUNCTIONAL TEST, IT WAS CONCLUDED THAT THE REPORTED "STAPLER FIRED WELL INITIALLY THEN FAILED TO FEED STAPLES" DURING SURGERY OCCURRED DUE TO A SEPARATED NOSE WELD IN INSTRUMENTS (B) AND (C) AND TO AN INCLINED STAPLE IN THE NOSE OF INSTRUMENT (B) AND A TWISTED STAPLE IN THE NOSE OF INSTRUMENT (B). NO CONCLUSION COULD BE REACHED AS TO IF THE STAPLES JAMMED IN THE NOSE OF THE DEVICES CAUSED THE NOSE WELD TO YIELD OR IF THE YIELDED NOSE WELDS CAUSED THE STAPLES TO BECOME JAMMED IN THE NOSE. INSTRUMENTS (A), (D), AND (E) WERE REC'D EMPTY IN GOOD CONDITION. INSTRUMENT (F) WAS CYCLED AND FIRED THE REMAINING 9 STAPLES WELL FORMED INTERMITTENTLY DUE TO INTERFERENCE WITH THE DRIED BODY FLUIDS IN THE CARTRIDGE NOSE.
IT WAS REPORTED DURING A LAPAROSCOPIC HERNIA REPAIR THE EMS STAPLER FIRED WELL INITIALLY THEN FAILED TO FEED STAPLES THEREAFTER (X6). 11/21/97 REP REPORTS SIX EMS STAPLERS WERE USED AND WITH EACH STAPLER THE FIRST STAPLE FIRED WITHOUT DIFFICULTY AND THEN JAMMED. A SEVENTH, NEW EMS WAS OPENED TO COMPLETE THE CASE. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ENDOSCOPIC MULTIFEED STAPLER | ENDOSCOPIC STAPLER | GAG | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |