FDA Adverse Event Malfunction Summary report: N

DREAMSTATION BIPAP AUTOSV

MDR report key: 13957184 · Received March 30, 2022

Report

Report Number
2518422-2022-06840
Event Type
Malfunction
Date Received
March 30, 2022
Date of Event
January 21, 2020
Report Date
April 13, 2022
Manufacturer
RESPIRONICS, INC
Product Code
MNS
PMA / PMN Number
K090539
Removal / Correction Number
RES 88058
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS MDR IS BEING SUBMITTED AS PART OF A BATCH SUBMISSION OF COMPLAINTS THAT WERE REASSESSED AS REPORTABLE FOAM DEGRADATION COMPLAINTS; DISCOVERED AS PART OF A RETROSPECTIVE REMEDIATION REVIEW.

Additional Manufacturer Narrative · 0

THE MANUFACTURER PREVIOUSLY REPORTED IN OUTCOMES ATTRIBUTED TO ADVERSE EVENT CONGENITAL ANOMALY/BIRTH DEFECT, BRAND NAME WAS REPORTED AS DREAMSTATION BIPAP AUTO, IT WAS REPORTED COMMON DEVICE NAME AS VENTILATOR, NON-CONTINUOUS(RESPIRATOR) AND PRODUCT CODE WAS REPORTED AS BZD AND PMA/ 510(K) NUMBER WAS REPORTED AS K131982 . AFTER REVIEW, IT WAS DETERMINED NO CONGENITAL ANOMALY/BIRTH DEFECT WAS REPORTED ,THIS WAS INCORRECTLY MARKED IN OUTCOMES ATTRIBUTED TO ADVERSE EVENT, BRAND NAME WAS MISTAKENLY REPORTED AS DREAMSTATION BIPAP AUTO., INCORRECTLY REPORTED COMMON DEVICE NAME AS VENTILATOR,NON-CONTINUOUS(RESPIRATOR) AND PRODUCT CODE WAS INCORRECTLY REPORTED AS BZD AND INCORRECTLY, PMA/ 510(K) NUMBER WAS REPORTED AS K131982. THE REPORT SHOULD HAVE BEEN SENT FOR PRODUCT PROBLEM ONLY IN ADVERSE EVENT OR PRODUCT PROBLEM. BRAND NAME WAS TAKEN AS DREAMSTATION BIPAP AUTOSV . DEVICE NAME WAS TAKEN AS VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING . AND PRODUCT CODE WAS TAKEN AS MNS. ADDRESS WAS UPDATED. PMA/ 510(K) NUMBER WAS TAKEN AS K090539. 'EVALUATION SUMMARY ATTACHED' WAS UN MARKED IN DEVICE EVALUATED BY MFR.

Description of Event or Problem · 0

THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP DEVICE'S SOUND ABATEMENT FOAM. THE DEVICE WAS RETURNED AND OBSERVE PARTICLE CONTAMINATION ON BOTTOM ENCLOSURE, TOP ENCLOSURE, AND ON REAR PANEL DURING DEVICE EVALUATION. THERE WAS NO ALLEGATION OF HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786884 DREAMSTATION BIPAP AUTOSV VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING MNS RESPIRONICS, INC DSX900T11C
843349 DREAMSTATION BIPAP AUTOSV VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING MNS RESPIRONICS, INC DSX900T11C

Patients

Seq Age Sex Outcome Treatment
1 Unknown Congenital Anomaly