FDA Adverse Event Malfunction Summary report: N

BD PHASEAL¿ INFUSION ADAPTER C100

MDR report key: 13953228 · Received March 30, 2022

Report

Report Number
3003152976-2022-00119
Event Type
Malfunction
Date Received
March 30, 2022
Date of Event
March 7, 2022
Report Date
April 28, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
ONB
UDI-DI
00382905153068
PMA / PMN Number
K123213
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: NO PHOTOS OR PHYSICALS SAMPLES THAT DISPLAY THE REPORTED ISSUE WERE PROVIDED FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2110233, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FIVE RETAINED SAMPLES OF THE REPORTED LOT WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCT WAS VISUALLY INSPECTED, THERE WAS NO DAMAGE OR OTHER DEFECTS OBSERVED ON ANY OF THE INFUSION ADAPTERS. PRODUCT UNDERGOES INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS TO ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE, INCLUDING VERIFICATION THAT ALL CRITICAL DIMENSIONS ARE WITHIN SPECIFICATION. ALL RETAINED SAMPLES UNDERWENT THESE EVALUATIONS AND PRODUCT WAS VERIFIED TO MEET REQUIRED LIMITS, INCLUDING PROPER SPIKE LENGTH AND DIAMETER. BASED ON THE AVAILABLE INFORMATION, WE CANNOT IDENTIFY A ROOT CAUSE RELATED TO OUR PRODUCT OR MANUFACTURING PROCESS AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED BD PHASEAL¿ INFUSION ADAPTER C100 SUFFERED A COMPONENT SEPARATION, RESULTING IN LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NURSE SPIKE BBRAUN INFUSION LINE INTO INFUSION ADAPTOR . WHEN BAXTER SOFT BAG WAS LIFTED UP , THE INFUSION ADAPTOR FELL OFF FROM THE BAG."

Description of Event or Problem · 0

IT WAS REPORTED BD PHASEAL¿ INFUSION ADAPTER C100 SUFFERED A COMPONENT SEPARATION, RESULTING IN LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NURSE SPIKE BBRAUN INFUSION LINE INTO INFUSION ADAPTOR. WHEN BAXTER SOFT BAG WAS LIFTED UP , THE INFUSION ADAPTOR FELL OFF FROM THE BAG."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1346038 BD PHASEAL¿ INFUSION ADAPTER C100 INTRAVASCULAR ADMINISTRATION SET ONB BECTON DICKINSON, S.A. 515306 2110233 00382905153068

Patients

Seq Age Sex Outcome Treatment
1 Unknown