FDA Adverse Event
Malfunction
Summary report: N
ALLEN MANUAL BEACH CHAIR
MDR report key: 1395267
·
Received March 11, 2009
Report
- Report Number
- 1221538-2009-00002
- Event Type
- Malfunction
- Date Received
- March 11, 2009
- Date of Event
- February 12, 2009
- Report Date
- March 11, 2009
- Manufacturer
- ALLEN MEDICAL SYSTEMS
- Product Code
- FWZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
DISASSEMBLY OF THE RETURNED CHAIR DEMONSTRATED THAT ONE OF THE SCREWS ATTACHED TO THE BACK OF THE UNIT WAS BROKEN. ONCE A WEIGHT/LOAD WAS PLACED IN THE CHAIR (SIMULATING A PT'S WEIGHT), THE BACK OF THE UNIT WAS NO LONGER ABLE TO HOLD POSITION AS INTENDED. ONCE THE SCREW WAS REPLACED AND THE CHAIR RE-ASSEMBLED, THE UNIT FUNCTIONED PROPERLY ONCE AGAIN.
Description of Event or Problem · 1
ON (B)(6)2009, AN ALLEN MEDICAL SALES REPRESENTATIVE WAS CONTACTED ABOUT AN ISSUE WITH THE ALLEN MANUAL BEACH CHAIR. THE BACK OF THE UNIT WAS NOT HOLDING POSITION PROPERLY WITH THE WEIGHT OF A PT IN IT. WITHOUT THE PT WEIGHT IN THE UNIT, THE DEVICE REPORTEDLY WAS WORKING FINE. THERE WERE NO INJURIES TO REPORT. THE FACILITY TOOK THE DEVICE OUT OF USE AND RETURNED IT TO US FOR EVAL AND REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALLEN MANUAL BEACH CHAIR | SHOULDER CHAIR, MANUAL BEACH CHAIR | FWZ | ALLEN MEDICAL SYSTEMS | O-SC-A3 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |