FDA Adverse Event Malfunction Summary report: N

ALLEN MANUAL BEACH CHAIR

MDR report key: 1395267 · Received March 11, 2009

Report

Report Number
1221538-2009-00002
Event Type
Malfunction
Date Received
March 11, 2009
Date of Event
February 12, 2009
Report Date
March 11, 2009
Manufacturer
ALLEN MEDICAL SYSTEMS
Product Code
FWZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DISASSEMBLY OF THE RETURNED CHAIR DEMONSTRATED THAT ONE OF THE SCREWS ATTACHED TO THE BACK OF THE UNIT WAS BROKEN. ONCE A WEIGHT/LOAD WAS PLACED IN THE CHAIR (SIMULATING A PT'S WEIGHT), THE BACK OF THE UNIT WAS NO LONGER ABLE TO HOLD POSITION AS INTENDED. ONCE THE SCREW WAS REPLACED AND THE CHAIR RE-ASSEMBLED, THE UNIT FUNCTIONED PROPERLY ONCE AGAIN.

Description of Event or Problem · 1

ON (B)(6)2009, AN ALLEN MEDICAL SALES REPRESENTATIVE WAS CONTACTED ABOUT AN ISSUE WITH THE ALLEN MANUAL BEACH CHAIR. THE BACK OF THE UNIT WAS NOT HOLDING POSITION PROPERLY WITH THE WEIGHT OF A PT IN IT. WITHOUT THE PT WEIGHT IN THE UNIT, THE DEVICE REPORTEDLY WAS WORKING FINE. THERE WERE NO INJURIES TO REPORT. THE FACILITY TOOK THE DEVICE OUT OF USE AND RETURNED IT TO US FOR EVAL AND REPAIR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLEN MANUAL BEACH CHAIR SHOULDER CHAIR, MANUAL BEACH CHAIR FWZ ALLEN MEDICAL SYSTEMS O-SC-A3 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention